A power fail condition that lasts longer than 10 seconds is also an RFS alarm and
will be indicated by the POWER FAIL LIGHTS red LED indicators and the beeper.
All RFS alarms are announced by an audible beeper that cannot be muted by pressing
the SILENCE ALARM button. The state of the RFS alarm is saved in the unit’s internal
memory and will not clear until the error is corrected and the RFS reset sequence is
performed.
Once the error condition has been corrected, the following RFS reset sequence needs to
be performed to clear the alarm from the equipment‘s internal memory:
Press and hold the COOL INCREMENT BUTTON and COOL TEMP SET
BUTTON and SILENCE ALARM buttons to clear the RFS alarm.
This will reset the RFS alarm, silence the audible beeper, and turn off the LED indicator.