OPERATION MANUAL
HEMOTHERM MODEL 400CE
Page
38
of
46
10. After the third rinse add to each reservoir the appropriate amount of distilled water per
the chart above and circulate.
11. Check the water with PH strips or other appropriate test method for Gigasept FF
detection. If Gigasept FF is detected, repeat Steps 8-10.
12.
RETURNING TO SERVICE:
If unit is to be
returned to service, replenish reservoirs
with the appropriate volume of distilled water. Proceed to Step 17.
13.
FOR DRY STORAGE:
If
unit is being
placed in dry storage, continue with procedure.
14. Drain the water from each reservoir as described in Section 3.4.
15. When all fluid has been removed from the unit, disconnect the drain hose(s) and wipe
unit clean.
16. Unit is now ready for storage. Refer to Section 1.11 for specifications for storage
conditions.
17. Document unit maintenance.
When you are ready to return unit to service, replenish the reservoirs with distilled water.
Perform all the checks as described in this Manual.
FLUID CIRCUIT CLEANING / DRY STORAGE PROCEDURE
For Circulating Water Units Utilizing CE Approved Maranon H (Product of Ecolab)
(Contact your local Cleaning Supplier)
(Maranon H Is Not Available In the United States)
The following procedure cleans the fluid circuit in these products. The required
tools/supplies are appropriate drain hose(s) and/or hose connector(s), Maranon H,
distilled water, and the appropriate AC electrical power
Procedure:
1.
Turn unit off (“O” position).
2. Remove any blood heat exchangers, Hyper-Hypothermia blankets, and hoses that are
connected to the unit.
3. Drain the water from the reservoir as described in Section 3.4.
4. Add to each reservoir the appropriate amount of water per the chart below:
RESERVOIR
MARANON H
WATER
TEMPERATURE
COOL
HEAT
230 Milliliters
170 Milliliters
7.6 Liters
5.7 Liters
15-25°C
37-42°C
5. Add to each reservoir the appropriate amount of Maranon H per the chart above.
6. Turn the unit on and circulate for five (5) minutes each in both cool and heat modes.
The unit utilizes an internal shunt to complete the circulation path.
7. Turn unit off.
8. Drain each reservoir as instructed in Section 3.4.
9. Rinse the unit three (3) times with water only as described in Steps 4-8.
Be sure not
to add the Maranon H.
10. After the third rinse add to each reservoir the appropriate amount of distilled water per
the chart above and circulate.
11. Check the water with PH strips or other appropriate test method for Maranon H
detection. If Maranon H is detected, repeat Steps 8-10.
Содержание HEMOTHERM 400CE
Страница 1: ......
Страница 47: ...HƯỚNG DẪN SỬ DỤNG MÁY TRAO ĐỔI NHIỆT HEMOTHERM MODEL 400CE Trang 47 46 ...
Страница 64: ...OPERATION MANUAL HEMOTHERM MODEL 400CE Page 17 of 46 FIGURE 1 2 KEYPAD 115V UNIT FIGURE 1 3 KEYPAD 115V REMOTE ...
Страница 65: ...OPERATION MANUAL HEMOTHERM MODEL 400CE Page 18 of 46 FIGURE 1 4 KEYPAD 230V UNIT FIGURE 1 5 KEYPAD 230V REMOTE ...
Страница 94: ...OPERATION MANUAL HEMOTHERM MODEL 400CE ...