OPERATION MANUAL
HEMOTHERM MODEL 400CE
Page
37
of
46
RESERVOIR
U.S.P. Grade Propylene Glycol Per Unit
COOL
HEAT
290 Milliliters
210 Milliliters
15.
Ensure that the water reservoirs are filled with distilled water. Proceed to Step 20.
16.
FOR DRY STORAGE:
If
unit is being
placed in dry storage, do the following steps.
17.
Drain the water from each reservoir as described in Section 3.4.
18.
When all fluid has been removed from the unit, disconnect the drain hose(s) and
wipe unit clean.
19.
Unit is now ready for storage. Refer to Section 1.11 for specifications for storage
conditions.
20.
Document unit maintenance.
When you are ready to return unit to service, replenish the reservoirs with distilled water.
Perform all the checks as described in this Manual.
FLUID CIRCUIT CLEANING / DRY STORAGE PROCEDURE
For Circulating Water Units Utilizing CE Approved Gigasept FF
(Product of
Schulke & Mayr
)
(Contact your local Cleaning Supplier)
(Gigasept FF Is Not Available In the United States)
The following procedure cleans the fluid circuit in these products. The required
tools/supplies are appropriate drain hose(s) and/or hose connector(s), Gigasept FF,
distilled water, and the appropriate AC electrical power
Procedure:
1.
Turn unit off (“O” position).
2. Remove any blood heat exchangers, Hyper-Hypothermia blankets, and hoses that are
connected to the unit.
3. Drain the water from each reservoir as described in Section 3.4.
4. Add to each reservoir the appropriate amount of water per the chart below.
RESERVOIR
GIGASEPT FF
WATER
TEMPERATURE
COOL
HEAT
800 Milliliters
600 Milliliters
7.6 Liters
5.7 Liters
15-25°C
37-42°C
5. Add to each reservoir the appropriate amount of Gigasept FF per the chart above.
6. Turn the unit on and circulate for fifteen (15) minutes each in both cool and heat
modes. The unit utilizes an internal shunt to complete the circulation path.
7. Turn unit off.
8. Drain each reservoir as instructed in Section 3.4.
9. Rinse the unit three (3) times with water only as described in Steps 4-8.
Be sure not
to add the Gigasept FF.
Reduce the water circulation time to five (5) minutes.
Содержание HEMOTHERM 400CE
Страница 1: ......
Страница 47: ...HƯỚNG DẪN SỬ DỤNG MÁY TRAO ĐỔI NHIỆT HEMOTHERM MODEL 400CE Trang 47 46 ...
Страница 64: ...OPERATION MANUAL HEMOTHERM MODEL 400CE Page 17 of 46 FIGURE 1 2 KEYPAD 115V UNIT FIGURE 1 3 KEYPAD 115V REMOTE ...
Страница 65: ...OPERATION MANUAL HEMOTHERM MODEL 400CE Page 18 of 46 FIGURE 1 4 KEYPAD 230V UNIT FIGURE 1 5 KEYPAD 230V REMOTE ...
Страница 94: ...OPERATION MANUAL HEMOTHERM MODEL 400CE ...