OPERATION MANUAL
HEMOTHERM MODEL 400CE
Page
5
of
46
WARNING
A physician's order is required for use and the setting of equipment and blanket
temperature. Check patient's temperature and skin condition of areas in contact with
blanket; also, check blanket water temperature at least every 20 minutes, or as directed by
physician. Pediatric patients, temperature-sensitive patients with vascular disease, surgical
patients, diabetics and Raynaud’s disease patients should be checked more frequently.
Notify the physician promptly of any change in patient status in order to avoid
serious injury or death.
The HEMOTHERM is to be operated by a professionally trained perfusionist
– with a
physician’s order for the setting of equipment and blanket temperature.
Do not use the HEMOTHERM MODEL 400CE Blankets distal to arterial cross clamping.
Non-observance can lead to thermal injury.
Thermal injury may occur
if heating/cooling blankets are used with a patient with an
ischemic limb.
Perfusionist must operate the HEMOTHERM when connected to extracorporeal circuit
heat exchanger.
Lack of proper monitoring may result in serious injury or death.
The operator of the unit regulates the desired fluid temperature.
Frequent observation
of the actual blood temperature and body temperature is required.
Perfusionist must monitor equipment continually during machine operation to ensure there
are no “alarms or intervention” required.
Lack of proper monitoring may result in
serious injury or death.
No modification of this equipment is allowed without prior, written authorization from CSZ.
Failure to do so may result in damage to the HEMOTHERM system and/or patient
injury
The method of temperature control provided by all cooler/heater units presents the danger
of heating or cooling body tissues, particularly the skin and or blood, to a point where they
are injured, i.e., burns or frostbite, respectively
. Depending on the extent and severity of
a burn, very serious and even fatal complications may arise.
Prevent excessive and/or prolonged tissue pressure and shearing forces, especially over
bony prominences.
Failure to do so may result in tissue injury.
Do not place additional heat sources between the patient and blanket.
Skin damage may
result.
Prep solutions have been reported to injure the skin when allowed to remain between
patients and a water-circulating heating blanket during prolonged procedures.
The area
between the patient and the blanket should be kept dry to avoid injury to patient.
Proper sanitation procedures must be practiced and hygienic safety must be
maintained, to prevent contamination.
Contamination can affect patient’s health, i.e.
skin irritation/rash or sepsis and infection may result.
Do not use the HEMOTHERM system in the presence of flammable anesthetics.
Risk of
explosion can result.
Power interruption will cause the HEMOTHERM to revert to
FILL MODE
resulting in no
therapy to the patient. Follow instructions for desired mode to resume operation.
Failure to
resume therapy could result in serious injury or death.
Any time water is found leaking into or around the unit, connecting hose, and/or blanket,
turn the unit off, disconnect the power cord from its power source, and correct the
problem before proceeding.
Water leaks could lead to electric shock. Water leaks
could present a slip and/or fall hazard.
Содержание HEMOTHERM 400CE
Страница 1: ......
Страница 47: ...HƯỚNG DẪN SỬ DỤNG MÁY TRAO ĐỔI NHIỆT HEMOTHERM MODEL 400CE Trang 47 46 ...
Страница 64: ...OPERATION MANUAL HEMOTHERM MODEL 400CE Page 17 of 46 FIGURE 1 2 KEYPAD 115V UNIT FIGURE 1 3 KEYPAD 115V REMOTE ...
Страница 65: ...OPERATION MANUAL HEMOTHERM MODEL 400CE Page 18 of 46 FIGURE 1 4 KEYPAD 230V UNIT FIGURE 1 5 KEYPAD 230V REMOTE ...
Страница 94: ...OPERATION MANUAL HEMOTHERM MODEL 400CE ...