4. Secondary High Limit Equipment Safety is a redundant system that is safety
preset at 46°C. If the device exceeds the maximum set-point of 42°C and the
primary high temperature safety should happen to fail at 44°C, a high temp light
on the key pad will illuminate and alarm at 46°C±0.5°C and the heaters will turn
off automatically. The circulating pump will no longer operate. If this happens, a
high temperature limit LED on the front panel will illuminate and the alarm will
sound.
The system should be inspected by the hospital's maintenance or
engineering department. This is a Remove from Service (RFS) alarm.
5. Circuit Breaker - The HEMOTHERM unit (right side) is equipped with an over-
current protection circuit breaker built into the power switch to protect the unit
wiring system against possible overload.