OPERATION MANUAL
HEMOTHERM MODEL 400CE
Page
20
of
46
14. HEAT DECREMENT BUTTON
– When pressed this button will decrease the
temperature set point indicated on the heat temperature display.
15. HEAT TEMP SET BUTTON
– When pressed this button will indicate the desired set
point temperature of the heat reservoir on the heat temperature display.
16. HEAT INCREMENT BUTTON
– When pressed this button will increase the temperature
set point indicated on the heat temperature display.
17. HEAT MODE BUTTON / LIGHT
– When pressed, this button will put the unit in heat
mode. It will start the heat reservoir conditioning to obtain the indicated set point
temperature. Unlike the cool mode button this button will allow the heaters to turn on
independent of the pump operation. The light will be illuminated when the unit is in heat
mode.
18. POWER FAIL LIGHTS
– When illuminated these lights will flash alternately to indicate
that during operation, the unit has become unplugged or otherwise lost power and the
power switch was in the on position.
19. HEAT LOW WATER LIGHT
– When illuminated water needs to be added to the heat
reservoir.
20. HEAT HIGH TEMP LIMIT LIGHT
– When illuminated the heat reservoir temperature is
higher than should be allowed by the safety limits.
1.4
FUNCTION OF HEMOTHERM MODEL 400CE COOLER/HEATER
The HEMOTHERM Model 400CE Cooler/Heater is used to lower/raise the blood
temperature and/or maintain blood temperature, as required, through conductive heat
transfer of conditioned water. The conductive heat transfer is accomplished through a
blood oxygenator/heat exchanger unit, along with the use of a blanket under the patient
for the re-warming phase, if desired. The complete system usually comprises the
HEMOTHERM Cooler/Heater unit and reusable or disposable blanket on the O. R. Table-
Adult, Pediatric, or Infant size. Available blankets are: the reusable PLASTIPAD
®
or
GELLI-ROLL
®
blankets (polyurethane) and the disposable MAXI-THERM
®
, and MAXI-
THERM
®
LITE blankets.
There is also an optional Remote Control, Model 414CE available that will allow the unit
to be controlled from 25 or 50 feet away.
Содержание HEMOTHERM 400CE
Страница 1: ......
Страница 47: ...HƯỚNG DẪN SỬ DỤNG MÁY TRAO ĐỔI NHIỆT HEMOTHERM MODEL 400CE Trang 47 46 ...
Страница 64: ...OPERATION MANUAL HEMOTHERM MODEL 400CE Page 17 of 46 FIGURE 1 2 KEYPAD 115V UNIT FIGURE 1 3 KEYPAD 115V REMOTE ...
Страница 65: ...OPERATION MANUAL HEMOTHERM MODEL 400CE Page 18 of 46 FIGURE 1 4 KEYPAD 230V UNIT FIGURE 1 5 KEYPAD 230V REMOTE ...
Страница 94: ...OPERATION MANUAL HEMOTHERM MODEL 400CE ...