INST.No.INE-567-P1CE
-28-
8.3 Data File and Folder
Remark
Caution
・
For one file, in the order from 1 to 9999 data can be saved.
・
Measured data is saved in a folder or a file set at setting mode.
・
Past data is not available to be overwritten.
・
If saving data exceeds 9999 data by continuous measurement etc, a new file will be created.
New file name will be "
○○○○
-01.csv" and forth until "
○○○○
-99.csv" saving is available
continuously. When saving data of "
○○○○
-99.csv" exceeds 9999 data, 'over' is displayed and
continuous saving becomes unavailable. Furthermore, data saved in "
○○○○
-
○○.csv
" is not
available to be replayed by the radiation thermometer. To replay the data, use PC etc.
・
If saved csv file is edited by PC etc., files may not be recognized by the thermometer.
・
Number of the file to be saved in the SD card is limited. So try not to save the files other than
measured data.
・
If data can not be saved in the SD card although there is enough free space in the card, it may
be a result of far too many files in the card. Delete some files or move some data to PC.
1 file
Folder
× Maximum 512 folders
9999 data
9999 files
[SD card (Maximum2GB)]