INST.No.INE-567-P1CE
-33-
9.5 Si Zero Adjustment (*IR-HAS ot IR-HAQ)
Performs Si zero adjustment.
Collimate the thermometer and sight the black body furnace to
input the temperature of the furnace.
1) From zero/span adjustment mode display, push
SEL
key several
times to display "Si Z" on the sub display.
2) Push
key or
key to change numeric value.
3) Blinking stops by pushing
ENT
key and the blinking digit moves
its position from lower place to the higher place and the
registration is completed.
9.6 Si Span Adjustment (*IR-HAS or IR-HAQ)
Performs Si span adjustment.
Collimate the thermometer and sight the black body furnace to
input the temperature of the furnace.
1) From zero/span adjustment mode display, push
SEL
key several
times to display "Si S" on the sub display.
2) Push
key or
key to change numeric value.
3) Blinking stops by pushing
ENT
key and the blinking digit moves
its position from lower place to the higher place and the
registration is completed.
Remark
Remark
Caution
Caution
At setting, if zero adjustment value (input temperature or measured temperature) exceeds span
adjustment value (input temperature or measured temperature), setting becomes disabled and
displays 'Er6'.
At setting, if span adjustment value (input temperature or measured temperature) is below span
adjustment value (input temperature or measured temperature), setting becomes disabled and
displays 'Er6'.
・
Setting range is 0 to 9999.
・
Setting of factory default for IR-HAS and IR-HAQ is both 600.
・
Setting range is 0 to 9999.
・
Setting of factory default for IR-HAS and IR-HAQ is 2000 and 3000.
[Si zero adjustment display]
Tb
CONT
MEM
PEAK
[Si
span adjustment display
]
Tb
CONT
MEM
PEAK