7. Measuring Mode
◆
There are two types of measuring method.
[Standard measuring mode] ······ This mode is for portable measurement.
[Continuous measuring mode] ··· This mode is for continuous measurement which this thermometer is
fixed and measuring for a long time at same location.
"
" under main marker "CONT" is lit.
Status flow is shown below.
For measuring method, refer to the individual sections.
Remark
Warning
Warning
If no key operation has done for 30sec. while in "HOLD" state, power supply is turned OFF
automatically.
Never directly sight the objective lens of the thermometer to the sunlight for protecting your eyes
and a detecting element.
・
For the measurement of an object exceeding 1500°C, make sure to turn the beam attenuation
filter switch "ON (attenuation side)" for protecting your eyes.
・
When you feel glare on the measurement of an objects lower than 1500°C, turn the beam
attenuation filter switch "ON (attenuation side)".
(Refer to "4.5 Cautions on Measurement")
◆
Setting Method
Start the thermometer by pushing
key
and push and hold
MEASURE
key while
power OFF.
*
under main marker "CONT" lights.
◆
Canceling Method
Start the thermometer by pushing
key
and push and hold
MEASURE key
while
power OFF.
*
under main marker "CONT" turns
OFF.
SEL
SEL
SEL
SEL
SEL
Flow Chart
Standard measuring
mode
Highest temperature
indication
Average temperature
indication
Emissivity
Thermocouple
temperature indication
Lowest temperature
indication
Continuous
measuring mode
[In the case thermocouple measurement is set
to "on"]
Push
SEL
key while "HOLD" status to display sub
marker TC.
Thermocouple measurement temperature data is
displayed on the sub display;
・
"oFL" is displayed at 1220°C or more
・
"uFL" is displayed at -50°C or less
*"oFL" is displayed if thermocouple is not
connected.
*Accuracy is not guaranteed for 800°C to
1220°C.