INST.No.INE-567-P1CE
-15-
<How to reset once set higher limit temperature alarm to "oFF">
1) Hold down
SEL
key to display "AL" on the sub display.
2) By changing displayed numeric value according to above
setting temperature
range with operation key (
key,
key or
ENT
key) to be below lower limit temperature
value, "oFF" is displayed.
(Lower limit temperature value is IR-HAI: 299°C or higher, IR-HAS: 599°C or higher, IR-HAQ:
399°C or higher)
3) By pushing
ENT
key in this status, blinking stops and "oFF" is displayed and registration is
completed.
6.2 Higher Limit Temperature Alarm
This setting is for a judgment of higher limit temperature alarm during measurement.
When the alarm is judged, status marker "AH" (higher temperature alarm) is lit and buzzer rings.
If "oFF" is selected, neither the alarm judgment nor the buzzer are activated.
1) From setting mode display, push
SEL
key several times to
display "AH" on the sub marker.
2) Push
key or
key to set OFF or numeric value.
3) Blinking stops by pushing
ENT
key and the blinking digit moves
its position from lower place to the higher place and the
registration is completed.
Reference
Caution
・
Setting temperature range is oFF or following
numeric values.
300 to 1000°C (IR-HAI) 600 to 2000°C (IR-HAS)
400 to 3000°C (IR-HAQ)
・
Default setting is "oFF".
Reference
[Higher limit temperature
alarm setting display]
Tb
CONT
MEM
PEAK