6.3 Signal Modulation Selection
This setting is used to extract nothing else than average value and maximum value continuously from the
measurement signal (original signal).
1) From setting mode display, push
SEL
key several times to
display "modu" on the sub display.
2) Push
key or
key to select signal modulation.
3) Blinking stops by pushing
ENT
key and the registration is completed.
dELy
It displays the value based on the first order lag of modulation degree set at "6.4.1 Modulation time
constant setting".
PEAk
It displays the value based on the original signal if temperature is rising.
If the temperature is dropping, the value set at attenuation rate setting (refer to the section 6.4.2) is
displayed.
6.4 Modulation Degree Setting
If "dELy" is selected in signal modulation selection, degree of first order lag can be adjusted by setting
modulation time constant.
If "PEAk" in signal modulation selection, signal attenuation degree after tracing peak can be set.
6.4.1 Modulation Time Constant
(
This is only valid if "dELy" is selected for signal modulation
)
1) From setting mode display, push
SEL
key several times to
display "tAu" on the sub display.
2) Numeric value on the main display blinks in the order of
0.0→0.2→0.5→1.0
(sec) when
key is pushed and
0.0→1.0→0.5→0.2
(sec) when
key is pushed.
3) Blinking stops by pushing
ENT
key and the registration is
completed.
Remark
Remark
Setting of factory default is "dELy".
Setting of factory default is "0.0"sec.
(Displayed value is based on the
original signal without any modulation).
[Modulation degree
setting display]
Tb
CONT
MEM
PEAK
[Signal modulation
selection display]
Tb
CONT
MEM
PEAK