3. Press the
key to edit the gain setting. The edit icon will be illuminated and
flashing.
4. Use the
and
keys to scroll through the GAIN settings in
PE/ PLAS (VLOW, LOW, MED, HIGH, VHI), or E-E, I-E, PLAS (LOW, MED,
HIGH) until the desired setting is being displayed.
NOTE:
Pressing the
key prior to pressing the
key will abort to the
measurement screen without saving changes.
5. Press the
key to set the GAIN and return to the measurement screen.
8.2 High Speed Scan
The High Speed Scan feature of the TI-007DLX increases the overall repetition rate to
a maximum of 140Hz with a high speed screen refresh rate of 25 times a second. This
allows for making scanned passes over an arbitrary length of the test material, while still
maintaining a reasonable representation of thickness over the area or region scanned.
The procedure to use the scan feature is outlined below:
1. Press the
key to access the menu items/features
2. Use the
and
keys to scroll through the menu items until the
SCAN feature is being displayed
3. Press the
key to edit the scan setting. The edit icon will be illuminated and
flashing.
4. Use the
and
keys to toggle scan on and off.
NOTE:
Pressing the
key prior to pressing the
key will abort to the
measurement screen without saving changes.
5. When the desired SCAN setting is displayed, press the
key to return to the
measurement screen.
s
8.3 Alarm
The Alarm feature of the TI-007DLX provides a method of setting tolerances, or limits,
for a particular application requirement. This feature may be used for a variety of
applications to verify the material thickness is within the manufacturer specifications.
The settings available are ON/OFF/BEEP, where beep enables the audible beeper.