– 25 –
1. Apply a drop of couplant on the transducer and place the transducer in steady
contact with the sample or actual test material. Be sure that the velocity
measurement is stable and the repeatability indicator, in the top left corner of the
display, is fully lit and stable.
2. Press the
key to enter the calibration edit screen displaying the current
velocity IN /μs (M /s) value. The edit icon will be illuminated and
flashing.
3. Use the
and
keys to scroll to the known velocity value.
NOTE:
Pressing the
key prior to pressing the
key will abort to the
measurement screen without saving changes.
4. Once the known velocity value is being displayed, press the
key to display
the calculated thickness based on known velocity.
5. Press the
key to return to the measurement screen and begin making
measurements.
NOTE:
The known velocity value that was entered will be displayed, and the thickness
value calculated will appear in the top right corner of the display for confirmation.