– 28 –
Both the on and beep settings will illuminate the led alarm lights above the keys on
the keypad. There are two limit values HI/LO, that can be set according to specified
tolerances.
The procedure to use the alarm feature is outlined below:
1. Press the
key to access the menu items/features.
2. Use the
and
keys to scroll through the items/features until the
ALRM feature is being displayed.
3. Press the
key to edit the alarm status. The edit icon will be illuminated and
flashing.
4. Use the
and
keys to toggle alarm on/off/beep.
5. When the desired ALRM status is displayed, press the
key to set the status
and edit the LO limit option.
6. Use the
and
keys to scroll the LO limit value to the target
value.
Note
: Pressing the
key prior to pressing the
key will abort to the
measurement screen without saving changes to the LO limit value.
7. When the target LO limit is displayed, press the
key to set the value and
advance to setting HI limit option.
8. Use the
and
keys to scroll the HI limit value to the target
value.
Note
: Pressing the
key prior to pressing the
key will abort to the
measurement screen without saving changes to the LO limit value.
9. When the target
HI
limit is displayed, press the
key to set the value and
return to the measurement screen.
8.4 Differential
The Differential Mode of the TI-007DLX provides the user with the ability to set a
nominal value, according to what the expected thickness should be, and measure the
+/- difference from the nominal value entered. This feature is typically used in QA
incoming inspections on pipes, plate stock, coils, etc.