10.2 electromagnetic immunity
recOMMeNdatiONS aNd declaratiON By tHe MaNuFacturer
cONcerNiNG electrOMaGNetic iMMuNity
Device is designed for use in the electromagnetic environment stipulated below. The buyer or user of
the device must ensure it is used in this recommended environment.
immunity test
test level iec
60601
Observance
level
electromagnetic environment -
recommendations
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV at the
contact
±8 kV in air
±6 kV at the
contact
±8 kV in air
Floors must be wood, concrete or
ceramic tile.
If floors are covered with synthetic
material the relative humidity must be
maintained at a minimum of 30%.
The Revolution Wireless device may be
susceptible to Electrostatic Discharge
(ESD) at and greater than ±15kV. In
the event of such a discharge device
may experience communication loss
between remote and modules. The
device will terminate all active outputs,
automatically place the device in a safe
state.
To recover from this, turn remote and
all modules off and turn them on after
5 sec delay using on/off switch on each.
Once the device restarts, re-initiate all
treatments.
Fast transient
electrical bursts
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
2 kV for power
supply lines
Not applicable
(I/O lines)
The quality of the electrical power
supply should be that of a typical
commercial or hospital environment.
Surge (1)
IEC 61000-4-5
±1 kV differential
mode
±2 kV joint mode
±1 kV differential
mode
Not applicable
(line to earth)
The quality of the power supply should
be that of a typical commercial or
hospital environment.