REVOLUTION WIRELESS™
daNGer!
Explosion Hazard
—
Revolution Wireless is not designed tor use in areas where
an explosion hazard may occur. An explosion hazard may result from the use of
flammable anesthetics, oxygen-rich environments, skin cleansing agents and
disinfectants.
Handle, clean, and dispose of components and accessories that have come
in contact with bodily fluids according to National, Local, and Facility rules,
regulations and procedures.
Stimulus delivered by the waveforms of this device, in certain configurations, will deliver
a charge of 25 microcoulombs (µC) or greater per pulse and may be sufficient to cause
electrocution. Electrical current of this magnitude must not flow through the thorax
because it may cause a cardiac arrhythmia.
Users with an implanted neurostimulation device must not be treated with or be in
close proximity to any shortwave diathermy, therapeutic ultrasound diathermy, or laser
diathermy anywhere on their body. Energy from diathermy (shortwave, microwave,
ultrasound, and laser) can be transferred through the implanted neurostimulation system,
can cause tissue damage, or can result in severe injury or death. Injury, damage, or death
can occur during diathermy therapy even if the implanted neurostimulation system is
turned off.
The solvents of adhesives and flammable solutions used for cleaning and disinfecting
should be allowed to evaporate before the unit is used.
DO NOT connect the unit to an electrical supply without first verifying that the power
supply is the correct voltage. Incorrect voltage may cause unit damage, malfunction,
electrical shock, fire, or personal injury. Your unit was constructed to operate only at the
elctrial voltage specified on the Voltage Rating and Serial Number plate. Contact your DJO
dealer if the unit is not properly rated.