1 0 . e M c ta B l e S
82
EN
REVOLUTION WIRELESS™
recOMMeNdatiONS aNd declaratiON By tHe MaNuFacturer
cONcerNiNG electrOMaGNetic iMMuNity
Device is designed for use in the electromagnetic environment stipulated below. The buyer or user of
the device must ensure it is used in this recommended environment.
immunity
test
test
level iec
60601
Observance
level
electromagnetic environment -
recommendations
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
150 kHz to 80
MHz
Portable and mobile RF communication devices must
only be used relative to the device and its wiring
at a distance which is not less than the spacing
recommended and calculated using the appropriate
equation for the transmitter’s frequency.
recommended spacing
d = 1.2 √
P
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
d = 1.2 √
P
80 MHz to 800 MHz
d = 2.3 √
P
800 MHz to 2.5 GHz
The Revolution Wireless device may be susceptible to
Radiated RF energy at and greater than 10V/m.
In the event of such exposure, the device may
experience communication loss between remote and
modules. The device will terminate all active outputs,
automatically place the device in a safe state.
To recover from this, turn remote and all modules
off and turn them on after 5 sec delay using on/off
switch on each. Once the device restarts, re-initiate
all treatments.
where
P
is the maximum output power of the
transmitter in watts (W) set by the manufacturer’s
specifications and where d is the recommended
spacing in metres (m). The field intensity of RF fixed
transmitters, as determined by an electromagnetic
survey a must be less than the observance level to
be found in each frequency range
b
. Interference
may occur close to any appliance identified by the
following symbol:
20xx
LATEX
FREE
REF