
60
EN
REVOLUTION WIRELESS™
6 . t r e at M e N t O P t i O N S , F u N c t i O N S a N d d e V i c e S e t t i N G S
PrOGraM NaMe
VMS-Fr dyNaMic 4 cH.
Program Description
The VMS™ FR Dynamic 4 Channels is a physiologically based channel
interaction in which one channel stimulates the agonist and the other
the antagonist of the muscle group that is being exercised. VMS is a
symmetrical biphasic waveform with a 100 μsec interphase interval.
Channels 1 and 2 will stimulate the first Agonist/Antagonist muscle
group while Channels 3 and 4 will stimulate the second Agonist/
Antagonist muscle group. Channels 1 and 2 are paired and Channels 3
and 4 are paired.
Waveform
VMS-FR
ParaMeterS
SettiNG
Treatment time
20 minutes
Burst Duration
2000 msec
Phase Duration
200µsec
Ramp Time
2 sec
Frequency
80pps
On/OFF Time
2 sec / 10 sec
Channels
1+2/3+4
Intensity range
1 to 120 mA
Muscle re-education