2 . S a F e t y i N F O r M at i O N
14
EN
REVOLUTION WIRELESS™
Warning!
equipment malfunction —
these warnings refer to potential equipment mailfunctions that
could result in patient hazards
− Magnetic and electrical fields are capable of interfering with the proper performance
of the unit. For this reason make sure that all external devices operated in the vicinity
of the unit comply with the relevant EMC requirements. ¬ Devices that intentionally
radiate electromagnetic energy, such as X-ray equipment, MRI devices, ¬radio systems
and cell phones are possible sources of interference, as well as RFID equipment (e.g.
electromagnetic security systems).
Keep the unit away from such equipment and verify its performance before use. For
additional information, refer to the EMC section of this Manual.
− Do not use the Revolution Wireless within 3 feet of short wave or microwave devices as
this could alter the currents generated by the stimulator. If you are in any doubt as to the
use of the stimulator in close proximity to another medical device, seek advice from the
manufacturer of the latter or from your doctor.
− Exercise caution when using electrotherapy while the patient is connected to monitoring
equipment with electrodes attached to the body. Stimulation could disrupt the signals
sent to the monitoring equipment.
− Refer
repair
and
maintenance
to authorized persons. Persons are authorized after
training by a specialist trained and commissioned by the manufacturer.
− Do not disassemble or modify the Revolution Wireless or its accessories. This may cause
damage, malfunction, electrical shock, fire, or personal injury.
− Inspect the
revolution
Wireless
and it’s accessories for damage and loose connections
at least
once a year
. Damaged and worn parts must be immediately replaced with
original spare parts by authorized staff.
- Do not use the device if you are connected to a high-frequency surgical instrument as
this could cause skin irritation or burns under the electrodes