79
EN
REVOLUTION WIRELESS™
1 0 . e M c ta B l e S
The Revolution Wireless needs special EMC precautions and must be installed and started according to the
EMC information supplied in this manual.
All RF wireless transmission systems can affect the Revolution Wireless.
The use of accessories, sensors and cables other than those recommended by the manufacturer may result
in stronger emissions or reduce the immunity of the Revolution Wireless.
The Revolution Wireless should not be used beside or stacked on top of any other equipment. If you must
use it side by side or on top of another system, you should check that the Revolution Wireless works
properly in the chosen configuration.
The product designation of the Revolution Wireless used in the text below includes all product variants.
This device is not intended to be used in a home environment
10.1 electromagnetic emissions
recOMMeNdatiONS aNd declaratiON By tHe
MaNuFacturer cONcerNiNG electrOMaGNetic eMiSSiONS
The device is intended for use in the electromagnetic environment specified below.
The customer or user of the device should ensure that it is used in this environment
emissions test
compliance
electromagnetic environment - Guide
RF emissions
CISPR 11
Group 1
The device uses RF energy only for its internal
operation. Consequently, its RF emissions are
unlikely to interfere with any adjacent electrical
device (radios, computers, telephones etc.).
RF emissions
CISPR 11
Class B
Device is suitable for use in any establishment,
other than a private dwelling or a place connected
directly to the low voltage mains supply which
powers residential buildings.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations / emission
oscillations IEC 61000-3-3
Complies