7
Intelect® Neo Clinical Therapy System
All modules (except vacuum module) are installed on
from the left side (when facing the screen) of the Neo
head unit and are each installed in the same manner.
Each has color-coded lead wires that correspond to the
appropriate colored labeling on the modules. Module-
speci
fi
c Installation instructions are shown after the
generic instructions. To install the modules in the Intelect®
Neo Clinical Therapy System, follow the steps shown.
NOTE:
Tools required (not included): #2 Phillips
screwdriver and standard slotted screwdriver.
The System is programmed to automatically recognize
the new Module(s), therefore, no software installation is
required.
Ultrasound
Laser
Stimulation 3/4
Stimulation 1/2
1. Ensure that the power cord is removed from the
device.
2. Remove the blank faceplate over the slot from the
left and right sides of the Neo head. (The example
displays the Ultrasound module.
3. Insert a standard slotted screwdriver (not provided)
into the top slot, pressing down with slight pressure.
Pull the faceplate off (in this example showing
electrotherapy channel 3/4).
4. The module is inserted on the left side of the Neo
Head in the slot as shown in this example (with the
Ultrasound module).
5. Carefully insert the module into the slot, with 32 pins
(2x16) in first. Secure the module in place with gentle
pressure until you feel the module is seated.
MODULE INSTALLATION
SETUP
Содержание Intelect Neo
Страница 1: ...Intelect Neo Clinical Therapy System User Manual Operator and Installation Instructions ...
Страница 2: ......