40
Intelect® Neo Clinical Therapy System
UTILITIES
6.
E-Stim Contact Quality
Select the <
E-Stim Contact Quality
> icon to turn on
or off the electrotherapy contact quality monitoring
function. The available choices are either “On” or “Off.”
The default setting is “On,” and in the “On” position,
the unit will monitor the contact quality of the
electrodes on all waveforms, excluding High Volt and
Microcurrent.
NOTE:
If the impedance feedback exceeds 5000
ohms (tol/-20%) and the set intensity is
≥
10 mA CC or 10 V CV, then the unit will display an
information message and set the intensity to zero for
all channels associated with this treatment.
7.
Restore Default Unit Settings
Select the <
Restore Default
> Unit Settings icon to
reset all of the following settings back to their factory
defaults:
• Volume
• LCD Brightness
• US Coupling
• US Head Warming
8.
Restore Default Protocols
Select the
<Restore Default Protocols>
icon to reset
all protocols (factory, custom, and the 5 custom
protocol shortcuts) to their factory defaults.
9.
Display Unit Version Information
Select the
<Display Unit Version Information>
icon to view the install status of each module. For
all installed modules, the PCB information and the
Software version for that module is displayed.
If a valid software upgrade USB flash drive is inserted
into the unit, the version of software for each module
that is on the USB flash drive is displayed.
Note: USB flash drive version is displayed in yellow if
it is newer than the current software version for the
applicable module.
10.
Export Protocols to USB flash drive
Select the Export Protocols to USB flash drive icon to
export all protocols (both default and custom) to a
valid USB flash drive.
UTILITIES AND OPTIONS
(
CONTINUED
)
11.
Import Protocols from USB flash drive
Select the Import Protocols from USB flash drive
icon to import all protocols from a valid USB flash
drive. Files exported from a differently configured
(e.g., Vectra® Neo vs Intelect® Neo) unit will not be
imported. Any protocols with the same name as
an existing unit protocol will be overwritten by the
imported protocol, and any protocols of a different
name on the unit will remain intact.
12.
Date and Time
Select the Date and Time icon to set the date and time
on the unit.
13.
Clinic Name
Select the <
Clinic Name
> icon to enter the name of
your clinic. The clinic displays on the Home Screen
and on the patient treatment reports saved to the USB
flash drive. The following buttons will perform the
functions listed:
•
Left
– Moves the cursor to the left one space per
instance pressed
•
Right
– Moves the cursor to the right one space per
instance pressed
•
Space
– Inserts one space per instance pressed
•
Clear
– Clears the entire entry
•
Delete
– Once the cursor has been moved using the
Left or Right icons, press the Delete icon to delete the
character
•
Caps
– Press the Caps icon to capitalize the letters or
to make letters lower case
•
Cancel
– Press the Cancel icon to cancel the clinic
name entry function and return to the Utilities screen
•
Accept
– Press the Accept icon once entry of the clinic
name is complete. The clinic name should now appear
in the title bar area of the screen