81
Intelect® Neo Clinical Therapy System
ELECTROMAGNETIC COMPATIBILITY
(
EMC
)
TABLES
Recommend separation distances between portable and mobile RF communications equipment and the
Intelect Neo Clinical Therapy System
The Intelect Neo Clinical Therapy System is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Intelect Neo Clinical Therapy System can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the Intelect Neo Clinical Therapy System as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = [3,5]
√
P
V1
80 MHz to 800 MHz
d = [3,5]
√
P
E1
800 MHz to 2,5 GHz
d = [7]
√
P
E1
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is a
ff
ected by absorption and re
fl
ection from structures,
objects and people.