IONTOPHORESIS PATIENT PREPARATION
PATIENT PREPARATION
• Appropriate skin preparation is required for adequate
conduction between the skin and the electrode. Clean
the skin with warm water. Do not use alcohol
• Choose the appropriate size electrode for the body
part (the larger the body part the larger the electrode
required)
• Larger electrodes are more comfortable and prevent
increases in current density
• Apply medicinal substance directly on skin (ointment)
or on gauze between skin and electrode, or inject in
iontophoresis electrode
• Attach the treating electrode over the treatment area
• Return electrode should be separated by at least the
diameter of the active electrode
• Dupel B.L.U.E. iontophoresis electrodes ordering
information:
- 199332-001 Small (1.5cc fill)
- 199334-001 Butterfly (2.0cc fill)
- 199335-001 Standard (2.5cc fill)
- 199336-001 Large (4.0cc fill)
- 1336 Black Snap Adapters (2 pin to snap adapters)
For Iontophoresis Operation, refer to page 55.