Intelect® Neo Clinical Therapy System
NOMENCLATURE
2. Press the Remote ON/OFF toggle icon to assign or
unassign the remote to the selected treatment.
The remote can be assigned to only one treatment
at a time however the remote can be reassigned as
needed.
When not in use, the Patient Remote/Laser Interrupt
Switch can be stored by hooking it onto the leadwire
holder clips in the same manner as leadwires and
cables, as demonstrated. Shown below.
• Disconnect the system from the power source before attempting
any maintenance, installation, removal or replacement procedures
to prevent electrical shock and possible damage to system.
• The laser interlock must be installed by a professional or qualified
electrician. Serious eye injury can result if the device is not
properly installed. Also, when installing the device for multiple
doors, the resistance total may not exceed 4800 ohm.
(")'%'*
INSTALLING THE LASER INTERLOCK (DOOR
INTERRUPT SWITCH)
The Laser Interlock is an optional safety device designed
to interrupt Laser therapy if the door to the therapy room
is opened. The laser interlock kit consists of a switch
resistor and a jack. Customers must supply the necessary
cable that complies with local and international codes.
Use only quali
fi
ed electricians to install the Laser Interlock
Kit.
Diagram for Therapy Room with One Door:
Diagram for Therapy Room with Multiple Doors: