background image

66

Intelect® Neo Clinical Therapy System

DATA

4.  View the selected image and if desired, print the 

selected image by pressing the printer icon:

5.  Press the Back arrow icon to scroll to the previous 

screen or Home icon to return to the Home screen.

ANATOMICAL LIBRARY

The Intelect® Neo Clinical Therapy System contains a 

unique Anatomical Library designed to aid the operator 

in visually understanding and locating speci

fi

 c muscle 

groups and commonly found problems associated 

with pathological conditions as well as providing an 

educational tool for the clinician to use with the patient. 

Complete the following steps to view the Anatomical 

Library:
1.  Press the Anatomical Library icon on the Home screen.

2.  Touch the body part for which you wish to view 

information. Choose either anterior (on left of screen) 

or posterior (on right of screen).

3.  Select an image from either the Anatomical Image list 

box or the Pathological Image list box by using the Up 

and Down arrows or Previous Page/Next Page icons 

the desired image. Select the image you wish to view 

and access by pressing the “

” symbol.

Содержание Intelect Neo

Страница 1: ...Intelect Neo Clinical Therapy System User Manual Operator and Installation Instructions ...

Страница 2: ......

Страница 3: ...1 LASER LABELS 32 WAVEFORMS 33 UTILITIES 39 UTILITIES AND OPTIONS 39 PATIENT PREPARATION 41 ELECTRODE PLACEMENT 41 DURA STICK ELECTRODES 41 ELECTROTHERAPY PATIENT PREPARATION 41 IONTOPHORESIS PATIENT PREPARATION 42 VACUUM ELECTRODE PATIENT PREPARATION 43 SEMG STIM PATIENT PREPARATION 45 LASER PATIENT PREPARATION 46 ULTRASOUND PATIENT PREPARATION 47 OPERATION 48 HOME SCREEN 48 TREATMENT SCREENS 49 ...

Страница 4: ...d module To maximize functionality and life of Intelect Neo be sure to Stay current with the latest clinical developments in the field of electrotherapy ultrasound laser therapy sEMG and sEMG electrotherapy Observe all applicable precautionary measures for treatment Keep informed of appropriate indications and contraindications for the use of the Intelect Neo Clinical Therapy System NOTE This equi...

Страница 5: ...virtually every therapeutic need that a clinician may have The components of the Intelect Neo Clinical Therapy System are shown below NOTE The Intelect Neo Clinical Therapy System when ordered as a Tabletop System without cart is assembled with Base as shown below The only user assembly required is the installation of the desired Modules described on page 7 Installation of the Vacuum module is req...

Страница 6: ...1 2 module the box will include the blue and green leadwires Stimulation Channel 3 4 is the cranberry and orange leadwires If both modules are ordered the box contains all four colored leadwires Stimulation modules channel 1 2 with sEMG includes blue and green sEMG leadwires Leadwire Holders Powercord COMPONENTS CONTINUED SETUP ...

Страница 7: ...ors used for the various modalities of the System Tools required not included 2 Phillips screwdriver and standard slotted screwdriver Remove the Intelect Neo Clinical Therapy System from the shipping carton Visually inspect for damage Report any damage to the carrier immediately To assemble the Intelect Neo Head to the Cart follow these steps 1 Remove the top drawer from the Cart Pull the drawer o...

Страница 8: ...n Press the plastic tabs on both drawer slides simultaneously in opposite directions as shown Completely pull the drawer out 2 There are two Cart height adjustments Standard shown on the left and lowered shown on the right For initial installation determine the desired height Locate three Allen style bolts for each leg left and right and insert by hand in their respective slots Use the Allen wrenc...

Страница 9: ...ly recognize the new Module s therefore no software installation is required Ultrasound Laser Stimulation 3 4 Stimulation 1 2 1 Ensure that the power cord is removed from the device 2 Remove the blank faceplate over the slot from the left and right sides of the Neo head The example displays the Ultrasound module 3 Insert a standard slotted screwdriver not provided into the top slot pressing down w...

Страница 10: ... 6 In this example showing the laser faceplate insert the faceplate at the bottom and snap into place at the top as shown on the left and right sides Neo allows Laser access on the left and right sides 7 Plug in the unit and press the power button allow the unit to initialize and then verify that the newly installed module is shown as available on the Home screen NOTE Installation of the Intelect ...

Страница 11: ...cm 2 in Round Disposable Electrodes 1 pack of 4 Faceplates to cover module after inserted into main unit Electrotherapy Module Channels 1 2 sEMG PN 70004 Stimulation module 2 channel Stimulation with sEMG sEMG Leadwires Dura Stick plus 5 cm 2 in 1 pack of 4 and Dura Stick plus 3 cm 1 25 in Round Disposable Electrodes Faceplates to cover module after inserted into main unit Laser Module PN 70005 La...

Страница 12: ...ower cord into the back of device Plug the other end of the cord into an electrical outlet NOTE The Power Cord may be unplugged from the back of the cart in an emergency situation 2 Press the Power button located on the top left portion of the LCD casing as shown below THERAPY SYSTEM START UP 3 Select desired function on the Home Screen shown below ...

Страница 13: ... Power Cord 10 Rear Access Panel 11 Serial Label 12 USB Flash Drive Port Flash Drive not included 13 Tilt Screen 14 Swivel function 15 Laser Interlock Port and Icon 16 Leadwire holders 6 7 13 45º Front Controls Side Holders 1 2 3 4 Rear Access Panel 14 5 6 7 8 10 9 11 12 16 15 NOMENCLATURE The Intelect Neo Clinical Therapy System Operator Interface contains all the functions and controls necessary...

Страница 14: ... Ultrasound or Combination therapy 2 Applicator The assembly that connects to the system and incorporates the Applicator 3 LED The component of the applicator that indicates if the Applicator is coupled or uncoupled on the treatment area 3 1 2 MODULE SLOTS ULTRASOUND APPLICATOR NOMENCLATURE To remove module take right side off face plate and push module from right side 1 Laser 2 Stimulation 1 2 St...

Страница 15: ... distributed by the applicator Pause Resume Button Laser Head LASER APPLICATOR LED s SLD s Laser Diode Aperture Laser Head NOTE When inserting the plugs for the Ultrasound and Laser modules be sure to align the flat side of the plug with the flat side of the slot and push in gently This is to avoid bending the pins in the plug Flat side of slot Flat side of plug ...

Страница 16: ...and resume if assigned to stim treatment M Manual Stimulation Electrical Stimulation treatments only When pressed for the first time will pause the treatment After the first press will ramp up and stimulate at desired output if held down or will pause treatment if released Complete the following steps to assign the remote to a treatment 1 When the remote is plugged into the unit a Remote icon is d...

Страница 17: ...res topreventelectricalshockandpossibledamagetosystem Thelaserinterlockmustbeinstalledbyaprofessionalorqualified electrician Seriouseyeinjurycanresultifthedeviceisnot properlyinstalled Also wheninstallingthedeviceformultiple doors theresistancetotalmaynotexceed4800ohm INSTALLING THE LASER INTERLOCK DOOR INTERRUPT SWITCH The Laser Interlock is an optional safety device designed to interrupt Laser t...

Страница 18: ...Neo Clinical Therapy System SYSTEM SOFTWARE SYMBOLS Stim Ultrasound Combo sEMG Laser CPS Custom Protocols Patient Data Anatomical Library Back Arrow Home Increase Decrease Parameter Scroll Up or Down in a text box Select Page up Page down Customize Save Data When pressed will print the screen contents or Patient Treatment Results Report to the USB Flash Drive Indicates a USB Flash Drive is Inserte...

Страница 19: ...en displays the following status information about each channel Not Installed Indicates the associated module is not installed in the unit Available Indicates the channel is available for use Setup Indicates a treatment for the channel is currently being setup but treatment has not yet begun Running Indicates a treatment for the channel is currently running Paused Indicates a treatment for the cha...

Страница 20: ...help DonotoperatethisunitwhenconnectedtoanyunitotherthanDJO devicesoraccessoriesspecificallydescribedinuserorservicemanuals Useofcontrols adjustmentsorperformanceofproceduresotherthan thosespecifiedhereinmayresultinhazardousexposuretoLaser energy DONOTdisassemble modify orremodeltheunitoraccessories This maycauseunitdamage malfunction electricalshock fire orpersonal injury Failuretouseandmaintaint...

Страница 21: ...gradedhydro gelcanresultininfectiontothe skinburn Useofelectrodeonmultiplepatientscanleadtoinfection PRECAUTIONARY INSTRUCTIONS Cleanapplicatorsaftereachuse otherwiseitcanleadtocross contaminationandinfection Cleanvacuumelectrodes spongesandhosesbeforeeachuse Lackof propercleaningandmaintenancemayleadtocross contamination andinfection Whenthelasermoduleisnotinuse itshouldbeprotectedagainst unquali...

Страница 22: ...Keepelectrodesseparatedduringtreatment Electrodesincontact witheachothercouldresultinimproperstimulationorskinburns Longtermeffectsofchronicelectricalstimulationareunknown Stimulationshouldnotbeappliedovertheanteriorneckormouth Severespasmofthelaryngealandpharyngealmusclesmayoccurand thecontractionsmaybestrongenoughtoclosetheairwayorcause difficultyinbreathing Stimulationshouldnotbeappliedtranstho...

Страница 23: ...y Carpal Tunnel Syndrome Epicondylitis Gouty arthritis Plantar Fasciitis Indications for FES Stimulation of the muscles of the leg and ankle of partially paralyzed patients to provide flexion of the foot and thus improve the patient s gait Contraindications The Intelect Neo Clinical Therapy System should NOT be used under the following conditions Do not use for symptomatic local pain relief unless...

Страница 24: ...sually be reduced by using an alternative conductive medium or an alternative electrode placement Electrode placement and stimulation settings should be based on the guidance of the prescribing practitioner Powered muscle stimulators should be used only with the lead wires and electrodes recommended for use by the manufacturer With TENS waveforms isolated cases of skin irritation may occur at the ...

Страница 25: ...en established for the use of therapeutic electrical stimulation during pregnancy Do not use powered muscle stimulators or TENS waveforms on patients with cardiac demand pacemakers Additional Precautions Use caution for patients with suspected or diagnosed heart problems Use caution for patients with suspected or diagnosed epilepsy Use caution in the presence of the following When there is a tende...

Страница 26: ...tissue necrosis might result Indications Application of therapeutic deep heat for the treatment of selected sub chronic and chronic medical conditions such as Relief of pain muscle spasms and joint contractures Relief of pain muscle spasms and joint contractures that may be associated with Adhesive capsulitis Bursitis with slight calcification Myositis Soft tissue injuries Shortened tendons due to...

Страница 27: ...form the following procedures to avoid the negative effects of Laser therapy Inspect the treatment area during and following treatment and discontinue if an adverse reaction does occur Higher output levels have a greater potential for patient discomfort Choose a lower dosage to reduce output or select a pulsed duty cycle to decrease patient discomfort Factors that Affect Treatment The following fa...

Страница 28: ...ttings 0 20 seconds in 1 second steps 0 5 seconds Power Input 20 25Vdc maximum peak current 4A Electrical Class CLASS I ElectricalType TYPE BF SYSTEM SPECIFICATIONS AND DIMENSIONS Width Depth Height Weight Vacuum 10 39 26 3906 cm 5 11 12 9794 cm 11 07 28 1178 cm 4 9lb 2 2226 kg Module 11 12 28 2448 cm 6 34 16 1036 cm 1 43 3 6322 cm 1lb 0 453592 kg Head 45 degree with Base Tabletop 15 89 40 3606 cm...

Страница 29: ... 1 MHz 0 10W at 3 3 MHz Medium Crystal 0 6W 1 and 3 3 MHz Small Crystal 0 3W 1 and 3 3 MHz Amplitude 0 to 2 5W cm2 in continuous mode 0 3W cm2 in pulsed modes Output accuracy 20 10 of maximum Temporal Peak to Average Ratio 2 1 20 at 50 Duty Cycle 5 1 20 at 20 Duty Cycle 9 1 20 at 10 Duty Cycle Beam Nonuniformity Ratio 6 1 maximum BeamType Collimating IPXX Rating for Unit IPX0 IPXX Rating for Appli...

Страница 30: ...ing the user that the probe is cooling down and the time period required After 5 seconds this message will disappear If the user attempts to use the probe before the cool down period is completed the message will re display to signify that the applicator is still in cool down mode After the cool down period is complete a message displays that informs the user that the unit is ready for use Single ...

Страница 31: ...W LED Five 850 nm 100 mW Laser 6 2218 0 097 0 54334 0 64 0 06 9 Diode Cluster Laser 1040 mW 1040 0 138 7 55 GaAIAs and LED Four 670 nm 10 mW LED Five 850 nm 200 mW Laser 12 4438 0 097 0 54334 0 64 0 188 13 Diode Cluster Laser 265 mW 265 0 035 7 55 GaAIAs and LED Seven 670 nm 10 mW LED Three 850 nm 50 mW Laser Three 950 nm 15 mW LED 3 1109 0 097 0 54334 1 12 0 136 0 48 13 Diode Cluster Laser 415 mW...

Страница 32: ...ED Twelve 670 nm 10 mW LED Thirteen 880 nm 25 mW LED Eight 950 nm 15 mW LED 0 386308 0 698132 N A 31 2 33 Diode Cluster Laser 690 mW 690 0 022 31 2 GaAIAs and LED Twelve 670 nm 10 mW LED Eight 880 nm 25 mW LED Eight 950 nm 15 mW LED Five 850 nm 50 mW Laser 3 1109 0 097 0 54334 1 92 1 28 1 28 06 33 Diode Cluster Laser 940 mW 940 0 030 31 2 GaAIAs and LED Twelve 670 nm 10 mW LED Eight 880 nm 25 mW L...

Страница 33: ...wavelength Each unit is shipped with laser protective eyewear that is L3 rated and approved as well as EN207 compliant Useful Range Optical Density 5 190 400 nm Optical Density 3 625 830 nm Optical Density 3 815 1050 nm Optical Density Wavelength SPECIFICATIONS LASER PROTECTIVE EYEWEAR SPECIFICATIONS ...

Страница 34: ...rial decal label is affixed at the rear of the system The device is labeled with the date of manufacturer The Laser applicator handle label includes important safety information The LED applicator handle label includes important safety information SPECIFICATIONS ...

Страница 35: ...alcurrentofthis magnitudemustnotflowthroughthethoraxbecauseit maycauseacardiacarrhythmia IFC Interferential Traditional 4 Pole Interferential Current is a medium frequency waveform Current is distributed through two channels four electrodes The currents cross each other in the body at the area requiring treatment The two currents interfere with each other at this crossing point resulting in a modu...

Страница 36: ... biphasic waveform with a 100 µsec interphase interval Because the pulse is relatively short the waveform has a low skin load making it suitable for applications requiring high intensities such as in muscle strengthening protocols Output Mode Electrodes Output Intensity 0 114 mA CC 0 56V CV Channel Mode Single Reciprocal Co Contract Phase Duration 20 400 µsec Mode Selection CC or CV Anti Fatigue O...

Страница 37: ...Hz 0 68V CV with carrier freq 5000 kHz Carrier Frequency 2 000 10 000 Hz Beat Fixed Sweep Off 1 200 Hz Sweep Low Beat Frequency 1 200 Hz Sweep High Beat Frequency 81 200 Hz CycleTime Continuous or User Defined Mode Selection CC or CV TreatmentTime 1 60 Min Available on Channel 1 2 3 or 4 Russian Russian Current is a sinusoidal waveform delivered in bursts or series of pulses This method was claime...

Страница 38: ...ut Mode Electrodes Output Intensity 0 72 mA Polarity Reversal On or Off With Polarity Reversal On Polarity will change every five minutes CycleTime Continuous 5 5 4 12 10 10 10 20 10 30 10 50 TreatmentTime 1 60 min Available on Channels 1 2 3 or 4 GALVANIC Interrupted Galvanic Current is a direct current flowing in one direction only The current can be continuous or interrupted Output Mode Electro...

Страница 39: ...hannel has a brief burst of power to slow down the initial acceleration of the agonist followed by a low output to regulate the movement of the agonist The movement is completed by a final burst of activity in both channels VMS is a symmetrical biphasic waveform with a 100 μsec interphase interval Because the pulse is relatively short the waveform has a low skin load making it suitable for applica...

Страница 40: ...ory chronic myofascial trigger points 4mg ml aqueous solution Glycopyrronium Bromide hyperhidrosis 0 05 solution Hyaluronidase edema scleroderma Reconstitute with 0 9 sodium chloride to provide a 150µg ml solution Hydrocortisone anti inflammatory 0 1 1 solution ointment Iodine sclerotic effect on scar tissue 5 10 solution ointment Lidocaine local anaesthetic effect 4 solution Lithium gouty arthrit...

Страница 41: ...s are available Off Nothing happens when applicator becomes uncoupled Pause and Beep When applicator becomes uncoupled treatment timer will pause unit will beep and channel area indicates uncoupled state Pause and No Beep When applicator becomes uncoupled treatment timer will pause unit will not beep and channel area indicates uncoupled state Beep When applicator becomes uncoupled treatment timer ...

Страница 42: ... the current software version for the applicable module 10 Export Protocols to USB flash drive Select the Export Protocols to USB flash drive icon to export all protocols both default and custom to a valid USB flash drive UTILITIES AND OPTIONS CONTINUED 11 Import Protocols from USB flash drive Select the Import Protocols from USB flash drive icon to import all protocols from a valid USB flash driv...

Страница 43: ... the skin for any wounds and clean the skin Apply the electrodes to the treatment area Ensure the electrodes are applied securely to the skin Ensure good contact between each electrode and the skin Check the electrode contact regularly during the treatment Examine the skin again after the treatment Choose electrodes that fit the anatomy View the Electrode Placement recommendations in the Treatment...

Страница 44: ...mfortable and prevent increases in current density Apply medicinal substance directly on skin ointment or on gauze between skin and electrode or inject in iontophoresis electrode Attach the treating electrode over the treatment area Return electrode should be separated by at least the diameter of the active electrode Dupel B L U E iontophoresis electrodes ordering information 199332 001 Small 1 5c...

Страница 45: ...ng treatment Connecting Lead Hoses 1 Insert the red lead hose WITH THE NARROW DIAMETER HOUSING to the positive lead hose connection port for the channel desired on the Vacuum Module 2 Connect the other lead hose again with the NARROW DIAMETER HOUSING to the negative lead hose connection port on the Vacuum Module NOTE Place the Vacuum plugs on all vacuum channels not being used to prevent vacuum le...

Страница 46: ...h the Vacuum Intensity Knob until the electrodes are secure on the surface Individually remove each electrode from the surface and place on the treatment area of the patient as prescribed NOTE Adjust Vacuum intensity until Vacuum Electrodes are secure to patient Note Do not use Vacuum Electrodes with sEMG sEMG STIM waveforms and Iontophoresis For Vacuum OPERATION refer to page 52 VACUUM ELECTRODE ...

Страница 47: ...ct sEMG 2 Choose one of the following modalities sEMG Ch 1 sEMG Ch 2 sEMG Ch 1 2 sEMG Stim VMS sEMG Stim Sym Biph Connect the sEMG Lead wires 1 Connect the sEMG Lead wires to the channel s desired for use with the selected modality connect Semg lead wires PATIENT PREPARATION 2 Select Customize 3 Select Electrode Placement to view electrode placement according to body area 4 Touch the body part you...

Страница 48: ...rimming or cutting electrodes may interfere with the reception of sEMG data and may affect the delivery of electrical stimulation in the sEMG Stim Modality 3 Review the specific Electrode Placement graphic for positioning of the Reference green lead electrode NOTE The electrodes may be placed for specific general and quasi specific biofeedback muscle or muscle group activity Intra Vaginal Probe If...

Страница 49: ...izes shown below Small Medium Large Applicator Preparation and Use 1 Clean applicator before each therapy session with warm soapy water 2 Liberally apply Conductor Transmission Gel to the treatment area on the patient Move the Applicator during therapy session in a circular motion The area treated should be twice the diameter of the Applicator 3 If US Coupling is On the Applicator is properly coup...

Страница 50: ...es access to all of the system modalities and functions The Home screen has the following information 1 Utilities Modality Icons 2 Electrotherapy 3 Ultrasound 4 Combo 5 sEMG 6 Laser Clinical Protocol Set up CPS Custom Protocols Assign Shortcut Options Patient Data Storage Anatomical Library OPERATION 1 2 3 4 5 6 ...

Страница 51: ...e Icon Press the Time icon to adjust therapy time duration 7 Intensity Icon Set intensity by rotating the Intensity Control Knob to the prescribed level Clockwise Increases Intensity Counterclockwise Decreases Intensity 8 Therapy Information Window View selected Therapy information such as Waveform Electrode Size CC CV Carrier Frequency Cycle Time Frequency and Vector Scan on the Therapy Informati...

Страница 52: ...ted in the same basic fashion The Intelect Neo Clinical Therapy System has an Electrotherapy Icon with the following waveforms IFC Interferential Traditional 4 Pole TENS Asymmetrical Biphasic TENS Symmetrical Biphasic TENS Alternating Rectangular TENS Monophasic Rectangular TENS HAN High Voltage Pulsed Current HVPC VMS Diadynamic Waveforms IFC Premodulated 2p Russian Microcurrent VMS Burst VMS FR ...

Страница 53: ...ode placement by dragging the electrodes to their appropriate locations on the body and selecting either front back left or right side Map pain information by touching the body area in the list box ELECTROTHERAPY OPERATION CONTINUED OPERATION Edit pain scale information by pressing the Edit Pain Scale icon Press the Up and Down arrows until pain is properly gauged Press the Back Arrow icon to retu...

Страница 54: ... below will then appear NOTE Combination Therapy sEMG and sEMG Electrical Stimulation Therapy cannot be administered using the Vacuum Electrode Module To view information explaining the waveform select the Modality Description icon Press the Up and Down icons to view additional text Press the Back Arrow icon to return to the previous screen or the Home icon to return to the Home screen Refer to th...

Страница 55: ...desired changes to the existing sequences by using the Up and Down arrows to find the desired sequence and press the Edit icon This screen also allows you to Delete a sequence from the protocol by pressing the Delete icon Add a new sequence to the protocol by pressing the New icon and using the Up and Down arrows to find the desired protocol in the list box Press the icon to add the new sequence S...

Страница 56: ... each time the Confirm icon is pressed If the waveform is Mono Triangular Pulsed then the phase duration will increase to next step each time the Confirm icon is pressed 5 Press the Start button to begin therapy the Pause button to pause treatment or the Stop button to terminate the treatment 6 When treatment has completed the Treatment Summary screen will appear with the following options Save th...

Страница 57: ...he treatment select the Pain Information icon At this screen you can do the following Map electrode placement by dragging the electrodes to their appropriate locations on the body and selecting either front back left or right side Map pain information by touching the body area in the list box Edit pain scale information by pressing the Edit Pain Scale icon Press the Up and Down arrows until pain i...

Страница 58: ...ons to view additional text Press the Back Arrow icon to return to the previous screen or the Home icon to return to the Home screen To view applicator information press the Applicator Information icon Touch the body part you wish to treat Information about applicator sizes frequency and treatment directions will appear as seen on the screen below Press the Down icon to scroll through the informat...

Страница 59: ...ion Combination therapy utilizes the Ultrasound modality in conjunction with High Voltage Pulsed Current HVPC IFC 4p IFC Premodulated 2p Asymmetrical Biphasic Symmetrical Biphasic or VMS to generate a therapeutic effect In this mode of therapy the Applicator of the Ultrasound Applicator becomes one half of the electrical circuit An electrode attached to the Black Lead Wire completes the circuit Co...

Страница 60: ...to set therapy intensity Clockwise Increases Intensity Counterclockwise Decreases Intensity 6 Press the Start button to begin therapy the Pause button to pause treatment or the Stop button to terminate the treatment 7 When treatment has completed the Treatment Summary screen will appear with the following options Save therapy information to USB Flash drive by inserting a USB Flash drive into the d...

Страница 61: ...ction Set up Steps To set the Target Threshold either Automatically or Manually touch the CUSTOMIZE Icon Adjust the Intensity with the Intensity Dial 5 Press Start Pause button to begin therapy Screen prompts will appear Contract Instructs the patient to attempt reaching the Target Threshold Contract appears on the screen indicating the patient should attempt to contract the selected muscle s Cont...

Страница 62: ...or saving customized settings 9 Press the Start button to begin therapy the Pause button to pause treatment or the Stop button to terminate the treatment 10 When treatment has completed the Treatment Summary screen will appear with the following options Save therapy information to USB Flash drive by inserting a USB Flash drive into the device and pressing the Save to USB Flash Drive icon Repeat th...

Страница 63: ... display To view information explaining the modality select the Modality Description icon Press the Up and Down icons to view additional text Press the Back Arrow icon to return to the previous screen or the Home icon to return to the Home screen 5 Verify that the Patient Remote Laser Interrupt Switch is connected and provided to the patient To view applicator information press the Applicator Info...

Страница 64: ...to USB Flash drive by inserting a USB Flash drive into the device and pressing the Save to USB Flash Drive icon Repeat the treatment by pressing the Run This Treatment icon Save the treatment protocol by pressing the Save Protocol icon Refer to the CUSTOM PROTOCOLS section for detailed instructions for saving customized settings LASER OPERATION CONTINUED OPERATION ...

Страница 65: ...e following options from the Treatment Summary screen Run the treatment by selecting the Run the Treatment icon Save the treatment protocol by pressing the Save Protocol icon Refer to the CUSTOM PROTOCOLS section for detailed instructions for saving customized settings View the pain information by selecting the View Pain Information icon Save therapy information to USB Flash drive by inserting a U...

Страница 66: ...the Save to USB Flash Drive icon Complete the following steps to delete a previously saved protocol 1 From the Home screen select the Custom Protocols icon 2 Find the previously saved protocol in the list box by using the Up and Down arrows or Previous Page Next Page icons to locate the protocol Select the protocol you wish to delete and press the Delete icon at the top of the screen as seen below...

Страница 67: ... to de assign protocol ______ shortcut 3 Select No to quit the deassignment process and return to the Home screen or Yes to continue with the deassignment process 4 After selecting Yes to delete the deassignment process a confirmation message will display for 5 seconds and return to the Home screen The previously assigned shortcut will no longer appear on the Home screen Complete the following ste...

Страница 68: ...mmonly found problems associated with pathological conditions as well as providing an educational tool for the clinician to use with the patient Complete the following steps to view the Anatomical Library 1 Press the Anatomical Library icon on the Home screen 2 Touch the body part for which you wish to view information Choose either anterior on left of screen or posterior on right of screen 3 Sele...

Страница 69: ...omplete existing treatment before attempting to start another treatment on the same channel 114 Information Attempting to select an sEMG treatment but no sEMG is installed on the Stim 1 2 module 1 If sEMG is installed on the Stim 1 2 module reset theTherapy System by turning theTherapy System Off then back On via the Power button 2 If sEMG is installed on the Stim 1 2 module and the problem persis...

Страница 70: ...ull Empty the vacuum reservoir as soon as possible 132 Information Attempting to perform a Print Screen to the USB flash drive while a treatment is running The Print Screen function is not allowed while a treatment is running Wait for treatment to complete and then try again 133 Information Attempting to import protocols from the USB flash drive while a treatment is running The Protocol Import fun...

Страница 71: ...or aTrainedTechnician before any further operation or use of the system Use of a system that indicates aWarning in these categories may pose a risk of injury to the patient user or extensive internal damage to the system Any 300 level code Error Immediately stop all use of the system and contact the dealer or DJO for service Errors in these categories indicate an internal problem with the system t...

Страница 72: ...cloth Round 40 Case 10 packs of 4 42197 5 cm 2 cloth Round 40 Case 10 packs of 4 42192 5 cm 2 foam Round 40 Case 10 packs of 4 42195 4 x 6 cm 1 5 x 2 5 foam Oval 40 Case 10 packs of 4 42196 5 x 10 cm 2 x 4 foam Oval 40 Case 10 packs of 4 42201 5 cm 2 cloth clip Square 40 Case 10 packs of 4 42202 5 x 10 cm 2 x 4 cloth clip Rectangle 40 Case 10 packs of 4 42204 5 cm 2 cloth Square 40 Case 10 packs o...

Страница 73: ...herapy Module Channels 3 4 70004 Electrotherapy Module Channels 1 2 sEMG 70005 Laser Module 70006 Vacuum Module 70008 PATIENT REMOTE LASER INTERRUPT SWITCH 79977 HIGHVOLT PROBE KIT Includes Probe and Sponge ApplicatorTips 15 and 8 mm 114 610exp Vaginal probe Saint Cloud Plus 114 502 Vaginal probe Saint Cloud Classic 114 242 Anal probe Saint Cloud w 2x2mm pin connectors 114 240 Anal probe Saint Clo...

Страница 74: ...Diode cluster 5 200mW laser diodes 4 LED s 27813 13 Diode cluster 3 50mW laser diodes 10 LED s 27814 13 Diode cluster 3 100mW laser diodes 10 LED s 27816 13 Diode cluster 3 200mW laser diodes 10 LED s 27802 33 Diode cluster 5 50mW laser diodes 28 LED s 27807 33 Diode cluster 5 100mW laser diodes 28 LED s 27808 33 Diode cluster 5 200mW laser diodes 28 LED s 27799 Single 670nm LED 10mW 27815 19 Diod...

Страница 75: ...n to complete 3 Enter utilities screen by pressing the Utilities button on the home screen 4 On the utilities screen press the Display Unit Version Information screen 5 Press the Upgrade Unit Software from USB button to apply any upgraded software packages to the unit COPY OF MANUAL To obtain a copy of the Intelect Neo Clinical Therapy System User Manual Item 13 7651 CD Version Item 13 7652 contac...

Страница 76: ...and Suction Cups A mild antibacterial solution containing no chlorine can be administered with a cloth and wiped or air dried This is recommended between patient treatments These electrodes are multiuse when properly maintained and cleaned Cleaning Instructions for the Sponges The associated sponges are recommended for single patient use only and should be cleaned with a 70 alcohol solution before...

Страница 77: ...inal invoice issued at purchase of theTherapy System or Module 3 Ship the unit to address specified by an authorized ServiceTechnician Service to these units should be performed only by a service technician certified by the Company Ultrasound Applicators require annual calibration from the date placed in service by the Factory or aTrainedTechnician Through the purchase of a Service Manual DJO LLC ...

Страница 78: ...Replacement parts or labor furnished by anyone other than the Company the selling dealer or a Company service technician Defects or damage caused by labor furnished by someone other than Company the selling dealer or a Company service technician Any malfunction or failure in the Product caused by product misuse including but not limited to the failure to provide reasonable and required maintenance...

Страница 79: ...istered This limit where the effect ceases is called the greatest active depth This depth is also contingent on tissue type pigmentation and dirt on the skin For example fat tissue is more transparent than muscle tissue and more easily penetrable by Laser than muscle tissue as blood hemoglobin is an important absorber of light and highly vascularized muscle tissue makes Laser penetration more diff...

Страница 80: ... and penetration of the Laser energy High powered diodes will reduce patients treatment times and give a higher amount of energy at a deeper depth Power output can be either continuous or pulsed Power Density Ratio of power divided by treatment time Pulsed Mode This is the ratio of the On time Total time of the cycle expressed as a percentage The lower the percentage the lower temporal average int...

Страница 81: ...ironment Guidance Electrostatic discharge ESD IEC 61000 4 2 6kV contact 8kV air 6kV contact 8kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines 1kV for input output lines 2kV for power supply lines 1kV for input output lines Mains pow...

Страница 82: ... transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 The...

Страница 83: ...r of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 3 5 P V1 80 MHz to 800 MHz d 3 5 P E1 800 MHz to 2 5 GHz d 7 P E1 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation di...

Страница 84: ...éen de Fret 3 rue de Bethar 64990 Mouguerre France T 33 0 5 59 52 86 90 F 33 0 5 59 52 86 91 http www djoglobal eu fr_FR index html 2013 DJO LLC All rights reserved Intelect Neo ClinicalTherapy System Manual 13 7651 Rev B 11 2013 ...

Отзывы: