67
Intelect® Neo Clinical Therapy System
Code
Number
Type Message Probable Cause
Possible Remedies
100
Information
Attempting to save a treatment to USB
fl
ash drive with a blank patient ID
Type in or select a Patient ID prior to saving treatment to USB
fl
ash drive
101
Information
Attempting to save a Custom protocol with a blank protocol name
Ensure protocol name is not blank prior to saving the protocol
102
Information
Attempting to save a Sequence with a blank sequence name
Ensure sequence name is not blank prior to saving the sequence
103
Information
Attempting to save the Clinic Name with a blank name
Ensure Clinic name is not blank prior to saving the Clinic name
104
Information
Attempting to delete a factory sequence
Factory sequences (names begin with a “*”) cannot be deleted
105
Information
Attempting to delete a factory protocol
Factory protocols (names begin with a “*”) cannot be deleted
106
Information
Attempting to save a Custom protocol or Sequence after system memory
has reached the maximum allowed (200)
Delete unused User protocols or sequences
107
108
109
Information
Information
Information
Attempting to access protocols or sequences and none are found in the
system
1. Custom protocols: No Custom protocols have been saved in the system. Refer to
the Custom protocols section
2. Sequences: No User sequences have been saved in the system. Refer to
Sequencing section
110
Information
Ultrasound applicator became unplugged during a treatment
Ensure ultrasound applicator is securely connected to the unit prior to starting an
ultrasound treatment
111
Information
No ultrasound applicator is plugged into the unit
Ensure desired ultrasound applicator is securely connected to the unit prior to
selecting an ultrasound treatment
112
Information
Attempting to select a treatment but available channels for desired
modality are currently in use
Complete existing treatment before attempting to start another treatment on the
same channel
113
Information
Attempting to select an sEMG treatment but sEMG channels are currently
in use
Complete existing treatment before attempting to start another treatment on the
same channel
114
Information
Attempting to select an sEMG treatment but no sEMG is installed on the
Stim 1/2 module.
1. If sEMG is installed on the Stim 1/2 module, reset the Therapy System by turning
the Therapy System O
ff
then back On via the Power button
2. If sEMG is installed on the Stim 1/2 module and the problem persists, discontinue
use of the device and contact the dealer or factory for technical service.
115
Information
Attempting to select a Laser treatment while another treatment is
running
Laser treatments cannot be run concurrently with another treatment. Complete
existing treatment(s) before attempting to select a laser treatment
116
Information
No laser applicator is plugged into the unit
Ensure desired laser applicator is securely connected to the unit prior to selecting a
laser treatment
117
Information
Laser applicator became unplugged during a laser treatment
Ensure desired laser applicator is securely connected to the unit prior to selecting a
laser treatment
118
Information
Incorrect Laser PIN was entered
Enter the correct laser PIN when prompted prior to starting a laser treatment
119
Information
Attempting to select a laser treatment but no laser module is installed in
the unit
1. If laser module is installed in the unit, ensure module is securely inserted
2. If laser module is not installed, contact the dealer or factory representative to
purchase one
120
Information
Attempting to start a Laser treatment but the remote control is not
plugged into the unit
Prior to starting a laser treatment, the remote control must be plugged into the unit
and given to the patient to allow use as an emergency stop
The Intelect® Neo Clinical Therapy System incorporates information, warning, and error messages to inform the user
of problems or potential problems with the system, modality, or accessories.
The numbering sequence is: 100-level
messages are general use information messages, 200-level messages are warnings, and 300-level messages are
errors.
Use the following Troubleshooting Chart to de
fi
ne the code and locate the probable cause and possible remedies
before contacting your dealer or the DJO o
ffi
ce.
TROUBLESHOOTING
TROUBLESHOOTING CODES