Advanced Settings (System Response Time)
The System Response Time is an adjustment that is used to make small adjustments to valve timing. It may be
required when the normal setup does not correctly dose bottles (or containers). This feature is intended to
compensate when the dosing head is greater than 3/4 inch from the bottle (or container) or if the pneumatic
pressure (GN
2
) is near the operational limits of the system. The initial value is set at the factory and does not
normally need to be adjusted.
1.
Use the arrow key
▲
or
▼
to go to the System Response Time screen (image below). Press ENTER to
highlight the value.
2.
Use the arrow key
▲
or
▼
to change the value. Press ENTER when the desired value is shown; range is
20 to 60 milliseconds (msec). The system response time is now set.