3 SCPI
3.3 Instrument-specific Commands
36
OFF|0: Its own ENR table is used for calibration when the status is disabled.
Example:
:SENSe:CORRection:ENR:COMMon:STATe OFF
:SENSe:CORRection:ENR:COMMon:STATe?
Reset status:
Common table on
Key path:
Front panel
【
ENR
】
—>[Common Table]
[ Auto Load ENR ]
:SENSe:CORRection:ENR:AUTO[:STATe]
Function description:
This command is used to set/query automatic loading status of the ENR table.
Setting format:
:SENSe:CORRection:ENR:AUTO[:STATe] OFF|ON|0|1
Query format:
:SENSe:CORRection:ENR:AUTO[:STATe]?
Parameter description:
ON|1: The measurement ENR table and related data are automatically loaded
from the SNS noise source when the status is enabled and one of the
following conditions is met:
When the SNS is connected for the first time;
The SNS is detected after it is powered on;
The SNS is connected when the command setting status is ON.
OFF|0: The status is off.
Example:
:SENS:CORR:ENR:AUTO:STAT OFF
:SENS:CORR:ENR:AUTO:STAT?
Reset status:
Common table on
Key path:
Front panel
【
ENR
】
—>[SNS Setup]—>[Auto Load ENR]
[ Calibration ENR Table ]
:SENSe:CORRection:ENR:CALibration:TABLe:DATA
Function description:
This command is used to set/query the data in the current calibration ENR table.
Setting format:
:SENSe:CORRection:ENR:CALibration:TABLe:DATA <freq>,<value>{,<freq>,<value>}
Query format:
:SENSe:CORRection:ENR:CALibration:TABLe:DATA?
Return value:
<freq>,<value>{,<freq>,<value>} The default units are Hz and dB
Parameter description:
<Freq> Frequency, the frequency units that can be input include Hz, kHz, MHz, GHz, and the
default unit is Hz
<value> ENR value in dB
Range: [1, 81] data groups
Example:
:SENSe:CORRection:ENR:CAL:TABL:DATA 1GHz,15.43dB,2GHz,15.2
:SENS:CORR:ENR:CAL:TABL:DATA?
Key path:
Front panel
【
ENR
】
—>[Common Table|Off]—>[Calibration ENR Table]—>[Edit]
Содержание 3986 Series
Страница 1: ...3986 Series Noise Figure Analyzer Programming Manual...
Страница 4: ......
Страница 8: ...3986 Series noise figure analyzer Table of Contents 4...
Страница 107: ...3 SCPI 3 3 Instrument specific Commands 99...
Страница 119: ...4 Programming example 4 2 Advanced operation example 111...