3 SCPI
3.3 Instrument-specific Commands
40
:SENSe:CORRection:ENR:SPOT?
Reset status:
15.2 dB
Key path:
Front panel
【
ENR
】
—>[ENR Mode Spot]—>[Spot]
[ Fixed ENR Mode ]
:SENSe:CORRection:SPOT:MODE
Function description:
This command is used to set/query the type of spot ENR and select the
parameter to be used during measurement.
Setting format:
:SENSe:CORRection:SPOT:MODE ENR|THOT
Query format:
:SENSe:CORRection:SPOT:MODE?
Return value:
ENR|THOT.
Parameter description:
The ENR value can be entered in Kelvin (K), Celsius (CEL), and Fahrenheit
(FAR), and the default unit is dB.
ENR: When this option is selected, it can be entered via the command
SENSe:CORRection:ENR:SPOT<value>;
THOT: When this option is selected, it can be entered via the command
SENSe:CORRection:ENR:THOT<value>.
Example:
:SENSe:CORRection:SPOT:MODE ENR
:SENSe:CORRection:SPOT:MODE?
Reset status:
ENR
Key path:
Front panel
【
ENR
】
—>[ENR Mode Spot]—>[Spot]—>[Spot Mode ENR|THot]
[ ENR THot]
:SENSe:CORRection:ENR:THOT
Function description:
This command is used to set/query the value of the fixed hot temperature when
the spot ENR is enabled.
Setting format:
:SENSe:CORRection:ENR:THOT <value>
Query format:
:SENSe:CORRection:ENR:THOT?
Return value:
Fixed hot temperature
Parameter description:
The ENR value can be entered in Kelvin (K), Celsius (CEL), and Fahrenheit
(FAR), and the default unit is K.
Example:
:SENSe:CORRection:ENR:THOT 16dB
:SENSe:CORRection:ENR:THOT?
Reset status:
9928.80 K (equivalent to a spot ENR of 15.2 dB)
Key path:
Front panel
【
ENR
】
—>[ENR Mode Spot]—>[Spot]—>[Spot Mode THot]—>[Fixed THot]
[ SNS TCold ]
:SENSe:CORRection:TCOLd:SNS[:STATe]
Function description:
This command is used to set/query the user cold temperature loaded from the
SNS.
Содержание 3986 Series
Страница 1: ...3986 Series Noise Figure Analyzer Programming Manual...
Страница 4: ......
Страница 8: ...3986 Series noise figure analyzer Table of Contents 4...
Страница 107: ...3 SCPI 3 3 Instrument specific Commands 99...
Страница 119: ...4 Programming example 4 2 Advanced operation example 111...