3 SCPI
3.3 Instrument-specific Commands
45
Query format:
:SENSe:CORRection:LOSS:AFTer :MODE?
Parameter description:
OFF: Loss compensation after DUT is off;
FIXed:The loss compensation value after DUT is a fixed value;
TABLe: The loss compensation value after DUT uses the data in the loss
compensation table.
Example:
:SENSe:CORRection:LOSS:AFTer:MODE FIXed
:SENSe:CORRection:LOSS:AFTer MODE?
Key path:
Front panel
【
Loss Comp
】
—>[After DUT]
Reset status:
Off
[ Loss Compensation Fixed Value after DUT ]
:SENSe:CORRection:LOSS:AFTer:VALue
Function description:
This command is used to set/query the loss compensation fixed value after DUT.
Setting format:
:SENSe:CORRection:LOSS:AFTer:VALue <value>
Query format:
:SENSe:CORRection:LOSS:AFTer:VALue?
Parameter description:
Range: [-100, 100] dB
Example:
:SENSe:CORRection:LOSS:AFTer:VALue 5dB
:SENSe:CORRection:LOSS:AFTer:VALue?
Key path:
Front panel
【
Loss Comp
】
—>[After DUT]—>[Fixed]—>[Fixed (After)]
Reset status:
0 dB
[ Loss Table Data after DUT ]
:SENSe:CORRection:LOSS:AFTer:TABLe:DATA
Function description:
This command is used to set/query the frequency-loss data pairs into the loss
compensation table after DUT.
Setting format:
:SENSe:CORRection:LOSS:AFTer:TABLe:DATA <freq>,<value>{,<freq>,<value>}
Query format:
:SENSe:CORRection:LOSS:AFTer:TABLe:DATA?
Parameter description:
201 groups of data can be input at most.
<freq> Frequency, which can be entered in Hz, kHz, MHz or GHz, and the default unit
returned by the query command is Hz
Range: [0, 100] GHz
<value> Loss value, which can be entered in dB.
Range: [-100, 100] dB
Example:
:SENSe:CORRection:LOSS:AFTer:TABLe:DATA 4.5GHz,5,6GHz,4
:SENSe:CORRection:LOSS:AFTer:TABLe:DATA?
Key path:
Front panel
【
Loss Comp
】
—>[After DUT]—>[Table]—>[Loss Table]—>[After Table]
Reset status:
the units are Hz and dB