3 SCPI
3.3 Instrument-specific Commands
98
Parameter description:
NORMal: Select normal noise source;
SNS: Select the smart noise source if it is connected; otherwise, use the normal
noise source.
Example:
:SENSe:SOURce:NOISe:STYLe NORMal
:SENSe:SOURce:NOISe:STYLe?
Reset status:
Smart
Key path:
Front panel
【
ENR
】
—>[SNS Setup]—>[Preference SNS|Normal]
3.3.16 System
[ GPIB Address ]
[:SENSe]:SYSTem:COMMunicate:GPIB:ADDRess
Function description:
This command is used to set/query the GPIB address of the noise figure analyzer.
Setting format:
[:SENSe]:SYSTem:COMMunicate:GPIB:ADDRess <integer>
Query format:
[:SENSe]:SYSTem:COMMunicate:GPIB:ADDRess?
Return value:
SNS|NORM
Parameter description:
<integer> GPIB address, integer, range: [1, 30]
Example:
:SYST:COMM:GPIB:ADDR 8
:SYST:COMM:GPIB:ADDR?
Reset status:
8
Key path:
Front panel
【
System/Local
】
—>[Interface Conf]—>[GPIB Addr]
3.3.17 Preset
[ Preset ]
:SYSTem:PRESet
Function description:
This command is used to reset the system.
Setting format:
:SYSTem:PRESet
Example:
:SYST:PRES
Key path:
Front panel
【
Preset
】
[ Preset ]
:DISPlay:PRESet
Function description:
This command is used to reset the system.
Setting format:
:DISPlay:PRESet
Example:
:DISP:PRES
Key path:
Front panel
【
Preset
】