3 SCPI
3.3 Instrument-specific Commands
60
[ Extension Module Set ]
:SENSe:CONFigure:EXTend:MODE
Function description:
This command is used to set/query the type of the extension module. The
instrument directly makes one-button measurement configuration and can carry
out extension measurements up to 110 GHz.
Setting format:
:SENSe:CONFigure:EXTend:MODE OFF|82411H|82411K|82411L|82411N|82411P
Query format:
:SENSe:CONFigure:EXTend:MODE?
Return value:
Extension module type
Parameter description:
Extension module type is optional,
OFF: The extension module is off;
82411H: Extension module frequency range: [50.0, 63.5] GHz
82411K: Extension module frequency range: [61.5, 75.0] GHz
82411L: Extension module frequency range: [75.0, 88.5] GHz
82411N: Extension module frequency range: [86.5, 100.0] GHz
82411P: Extension module frequency range: [96.5, 110.0] GHz
Example:
:SENS:CONF:EXT:MODE 82411K
:SENS:CONF:EXT:MODE?
Reset status:
OFF
Key path:
Front panel
【
Mode Setup
】
—>[Extend Setup]
3.3.6
Calibration
The calibration corresponds to the
【
Calibrate
】
key on the front panel. This key is used to perform
measurement calibration. The value generated during calibration is used for the corrected
measurement.
[ Calibration ]
:CALibration
Function description:
This command is used to perform the calibration operation in the instrument.
Setting format:
:CALibration
Example:
:CAL
Key path:
Front panel
【
Calibrate
】
Description:
The calibration means that the sweep measurement is performed with the frequency
step amount defined by the frequency points from the start frequency to the stop
frequency.
[ Calibration ]
:CALibration:AUTO:CALibration
Function description:
This command is used to perform the calibration operation in the instrument.
Setting format:
:CALibration:AUTO:CALibration OFF|ON|0|1
Parameter description:
ON|1
:
The calibration is on
OFF:|0: The calibration is off