3 SCPI
3.3 Instrument-specific Commands
47
Key path:
Front panel
【
Mode Setup
】
—>[DUT Setup]—>[DUT]
Reset status:
Amplifier
[ DUT LO Type ]
:SENSe:CONFigure:MODE:DUT:LOSCillator
Function description:
This command is used to set/query the status of the LO when the device under
test is measured by the converter: LO fixed or LO variable.
Setting format:
:SENSe:CONFigure:MODE:DUT:LOSCillator FIXed|VARiable
Query format:
:SENSe:CONFigure:MODE:DUT:LOSCillator?
Parameter description:
FIXed: LO frequency remains a fixed constant;
VARiable: The LO frequency is variable.
Example:
:SENSe:CONFigure:MODE:DUT:LOSCillator FIXed
:SENSe:CONFigure:MODE:DUT:LOSCillator?
Reset status:
Fixed
Key path:
Front panel
【
Mode Setup
】
—>[DUT Setup]—>[LO Mode]
Description:
Note that LO fixed means that the IF frequency is variable and LO variable
means that the IF frequency is fixed.
This command is only used when measuring up or downconverter devices or
system down converter.
[ System DownConverter Control ]
:SENSe:CONFigure:MODE:SYSTem:DOWNconv[:STATe]
Function description:
This command is used to set/query the measurement status of the system down
converter.
Setting format:
:SENSe:CONFigure:MODE:SYSTem:DOWNconv[:STATe] OFF|ON|0|1
Query format:
:SENSe:CONFigure:MODE:SYSTem:DOWNconv[:STATe]?
Parameter description:
ON|1: System downconverter on
OFF|0: System down converter off
Example:
:SENSe:CONFigure:MODE:SYSTem:DOWNconv OFF
:SENSe:CONFigure:MODE:SYSTem:DOWNconv?
Reset status:
Off
Key path:
Front panel
【
Mode Setup
】
—>[DUT Setup]—>[DownConv]
[ System DownConverter LO Mode ]
:SENSe:CONFigure:MODE:SYSTem:LOSCillator
Function description:
This command is used to set/query the system downconverter LO mode: fixed or
variable.
Setting format:
:SENSe:CONFigure:MODE:SYSTem:LOSCillator FIXed|VARiable
Query format:
:SENSe:CONFigure:MODE:SYSTem:LOSCillator?
Parameter description:
FIXed: LO frequency remains a fixed constant;
Содержание 3986 Series