3 SCPI
3.3 Instrument-specific Commands
34
[ Average ]
:SENSe:AVERage:COUNt
Function description:
This command is used to set/query the number of samples per measurement
during averaging.
Setting format:
:SENSe:AVERage:COUNt <integer>
Query format:
:SENSe:AVERage:COUNt?
Parameter description:
<integer> Integer
Range:[1, 999].
Example:
:SENS:AVER:COUN 20
Reset status:
8
Key path:
Front panel
【
Average
】
—>[Average]
[ Average Mode ]
:SENSe:AVERage:MODE
Function description:
This command is used to set/query the average calculation method. The
averaging method is point or sweep average.
Setting format:
:SENSe:AVERage:MODE POINt|SWEep
Query format:
:SENSe:AVERage:MODE?
Return value:
POIN Point average
SWE Sweep average
Parameter description:
POINt: during the sweep, averaging is performed at each point based on the
set number of average before the measurement is carried out at the
next point. The measurement is completed at the end of one sweep.
SWEep: Each point is averaged once during the sweep. The result of each
point is obtained after multiple sweep. The number of sweeps is the
set number of average.
Example:
:SENSe:AVERage:MODE POINt
:SENSe:AVERage:MODE?
Reset status:
POINt
[ Measurement Bandwidth ]
:SENSe:BANDwidth[:RESolution]
Function description:
This command is used to set/query the measurement bandwidth.
Setting format:
:SENSe:BANDwidth[:RESolution] 4Mhz|2Mhz|1Mhz|400Khz|200Khz|100Khz
Query format:
:SENSe:BANDwidth[:RESolution]?
Parameter description:
Fixed frequency value:
4Mhz|2Mhz|1Mhz|400Khz|200Khz|100Khz
Example:
:SENS:BAND 2MHz
:SENS:BAND?