3 SCPI
3.3 Instrument-specific Commands
84
Key path:
Front panel
【
Correct
】
—>[RF Max Cal Att]
[ Microwave Minimum Attenuation ]
:INPut:ATTenuation:MWAVe:MINimum
Function description:
This command is used to set/query the microwave minimum attenuation when
the calibration is performed.
Setting format:
:INPut:ATTenuation:MWAVe:MINimum <integer>
Query format:
:INPut:ATTenuation:MWAVe:MINimum?
Parameter description:
<integer> Range: [0, 10] dB, with the step of 5 dB
Return value:
Microwave minimum attenuation when the calibration is made
Example:
:INPut:ATTenuation:MWAVe:MINimum 5dB
:INPut:ATTenuation:MWAVe:MINimum?
Reset status:
0 dB
Key path:
Front panel
【
Correct
】
—>[uW Min Cali Attn]
[ Microwave Maximum Attenuation ]
:INPut:ATTenuation:MWAVe:MAXimum
Function description:
This command is used to set/query the microwave maximum attenuation when
the calibration is performed.
Setting format:
:INPut: ATTenuation:MWAVe:MAXimum <integer>
Query format:
:INPut:ATTenuation:MWAVe:MAXimum?
Parameter description:
<integer> Range: [10, 15] dB, with the step of 5 dB
Return value:
Microwave maximum attenuation when the calibration is made
Example:
:INPut:ATTenuation:MWAVe:MAX 15dB
:INPut:ATTenuation:MWAVe:MAX?
Reset status:
10 dB
Key path:
Front panel
【
Correct
】
—>[uW Max Cal Att]
3.3.12 Sweep
The sweep corresponds to the
【
Sweep
】
key on the front panel, including sweep mode, manual
measurement and other commands.
3.3.12.1 Sweep Control
[ Sweep Mode ]
:INITiate:CONTinuous
Function description:
This command is used to set/query whether the continuous sweep starts.
Setting format:
:INITiate:CONTinuous OFF|ON|0|1
Query format:
:INITiate:CONTinuous?