3 SCPI
3.3 Instrument-specific Commands
46
[ Number of Input Loss Table after DUT]
:SENSe:CORRection:LOSS:AFTer:TABLe:COUNt?
Function description:
This command is used to query the number of input loss compensation after
DUT.
Query format:
:SENSe:CORRection:LOSS:AFTer:TABLe:COUNt?
Parameter description:
Range: [0, 201] dB
Example:
:SENSe:CORRection:LOSS:AFTer:TABLe:COUNt?
[ Loss Compensation Temperature after DUT ]
:SENSe:CORRection:TEMPerature:AFTer
Function description:
This command is used to manage the temperature of the loss compensation
after DUT.
Setting format:
:SENSe:CORRection:TEMPerature:AFTer <temperature>
Query format:
:SENSe:CORRection:TEMPerature:AFTer?
Parameter description:
The unit is Kelvin (K), Celsius (CEL), Fahrenheit (FAR) or K, C, F. The default
unit returned by the query is K.
Range: [0, 29650000.00] K
Example:
:SENSe:CORRection:TEMPerature:AFTer 1000K
:SENSe:CORRection:TEMPerature:AFTer?
Key path:
Front panel
【
Loss Comp
】
—>[Temp (after)]
Reset status:
290.0K
3.3.5 Mode Setup
The mode set corresponds to the
【
Mode Setup
】
key on the front panel, which is used for selecting the
type of device under test, setting mode measurement, configuring external local oscillator, calculating
uncertainty, and setting extension module.
3.3.5.1 DUT Setup
The DUT set corresponds to the
【
Mode Setup
】
[DUT Setup] key on the front panel, which is used to
select the type of device under test and the measurement settings under the corresponding type,
including DUT type, LO type fixed or variable, system down converter control switch, fixed IF frequency,
fixed LO frequency, LO frequency offset setting, frequency mode, and input frequency properties.
[ Select DUT Type ]
:SENSe:CONFigure:MODE:DUT
Function description:
This command is used to set/query the type of the device under test.
Setting format:
:SENSe:CONFigure:MODE:DUT AMPLifier|DOWNconv|UPConv
Query format:
:SENSe:CONFigure:MODE:DUT?
Parameter description:
AMPLifier: The measured DUT is of the amplifier type;
DOWNconv: The measured DUT is a downconverter component;
UPConv: The measured DUT is an upconverter component.
Example:
:SENS:CONF:MODE:DUT AMPL
:SENS:CONF:MODE:DUT?