LSS-2404
24 |
P a g e
Test Defaults (1)
Used to transfer default values into RAM, also setting them to be the Current settings. The values should
immediately appear in the GUI. Since default trigger type = one-shot, it waits for trigger before executing.
Note: This will change the recipe settings but not the IP address
Test (2)
Used to transfer current GUI settings into RAM, i.e. it establishes the Active settings. If trigger type = free run,
the recipe runs for testing; otherwise it waits for a trigger.
Also, it stops a sequence in the middle of its execution if pressed while the sequence is running, and trigger
event is set to One-Shot or Continuous
Display Current (3)
Displays the Active settings stored in RAM.
Load from Controller (4)
Loads and displays the saved settings stored in flash. It will overwrite the Active settings in RAM.
Save to Controller (5)
Saves the GUI settings to flash and overwrites the Active settings in RAM. After click
ing ‘S
ave
to Controller’
button, see message displayed that Save is complete.
Save to File (6)
Saves the GUI settings to named file and overwrites the Active settings in RAM. After clicking
‘
Save
to File’
button, see message displayed that Save is complete. It will not change flash settings.
Note: The file needs to be in the same directory as the LSS2404_GUI.html
Load from File (7)
Loads and displays the saved settings stored in the selected file. It will overwrite the Active settings in RAM. It
will not change flash settings. Use Browse to select the file and Load from File to upload to LSS GUI.
Help (8)
Loads a document to assist with programming the GUI
GUI IP Address (9)
Used to distinguish which LSS box you are programming if using multiple LSS units in a network. Type in the IP
address of the LSS box being programmed and the GUI will connect to that box.
Update Firmware (11)
Updates the LSS controller to the selected firmware. See ‘Update Firmware’ above for instructions on how to
update.