27
CAL 6(Zero-set adjustment)
Function: Zero adjustment - calibration when any zeroing error occurs.
Used key
Display
Descriptions
= Set,
=Cancel
2-CAL
Empty the load tray and press the setup
key.
1234
The current weight value is displayed.
Confirm 'Stable' and press the setup key.
- - -
Zero adjustment in progress...
Reference 1. Use when the zeroing is failing due to shock. (Zero-set range is 0 ~ 2 mV/V)
Reference 2. If zero-set adjustment is completed without error, advance to 'CAL-1'.
CAL 7(Factor Calibration)
Function: Factor Calibration
Used key
Display
Descriptions
Using numeric keys
~
Set the weight factor.
= Set,
=Cancel
Push
when entering a
decimal print
NOTUSE
This function cannot be used because
of multi calibration.
FACtor
It means you entered the factor
correction mode.
12345
The current factor is displayed.
Reference 1. This menu is for weight calibration when counterweight is not available. Regular users will not need to
use this.
Reference 2. Can only be used when the multicalibration range of 'CAL3-1' is set to 1.
Reference 3. To enter the FACTOR adjustment mode, PASSWORD must be entered.
Содержание CI-400 Series
Страница 1: ... 2 ...
Страница 10: ...11 2 Specifications in Appearance 2 1 External Dimension ...
Страница 29: ... 30 4 2 How to Seal the Indicator Sealing ...
Страница 93: ... 94 Connection of Input Output Input AContact Output AContact ...
Страница 104: ...105 Weight Date 8 byte Example weight Byte No 1 2 3 4 5 6 7 8 13 5kg 1 3 5 135kg 1 3 5 135kg 1 3 5 ...
Страница 117: ... 118 MEMO ...
Страница 118: ...119 MEMO ...
Страница 119: ... 120 MEMO ...
Страница 120: ...121 ...