31
5. Weighing Mode
(1) Zero-set Function (used when changing the zero-set)
Display Part or Used Keys
Load Plate
Description
Step 1
Empty
State with zero changed
Step 2
Push the zero key
Step 3
Empty
State after performing zero
function.
Namely, the current weight is
designated as '0'kg.
Note 1. Operating range for the zero key is possible between ±1% ~±99% of the maximum
weight. Operating range for the zero key is designated in Menu No. [F1.15].
Note 2. Menu No. [F1.14] designates whether to perform zero function only if the current
weight is stabilized or even when it is unstable.
(2) Tare function (used when weighing using a container)
■
Maximum tare setting range: maximum capacity
Display Part or Used Keys
Load Plate
Description
Step 1
Tare Placement
State with tare
placed on load plate
Weight plate : 1.000 kg
Step 2
Push the tare key
Step 3
Tare
State with NET lamp turned
ON and tare registered
Содержание CI-400 Series
Страница 1: ... 2 ...
Страница 10: ...11 2 Specifications in Appearance 2 1 External Dimension ...
Страница 29: ... 30 4 2 How to Seal the Indicator Sealing ...
Страница 93: ... 94 Connection of Input Output Input AContact Output AContact ...
Страница 104: ...105 Weight Date 8 byte Example weight Byte No 1 2 3 4 5 6 7 8 13 5kg 1 3 5 135kg 1 3 5 135kg 1 3 5 ...
Страница 117: ... 118 MEMO ...
Страница 118: ...119 MEMO ...
Страница 119: ... 120 MEMO ...
Страница 120: ...121 ...