<Limit mode 2> Relay Operation Graph upon Setting (F6.03-1)
1. Set value input requirement : (SP2 – SP4) > (SP1 – SP3)
2. T1 : Refer to F6.04 (Delay time of weighing Finish relay output)
T2 : Refer to F6.05 (Operation time of weighing Finish relay output)
T5 : Refer to F6.08 (Operation(ON) time of Weighing NG relay output)
3. Relay Output
Step1 : ON when (SP1 – SP3 ) ≤
Weight
Step2 : ON when (SP2 – SP4 ) ≤
Weight
Finish : On after T1 (set time), ON after T2 (set time)
Lowest Limit NG: ON when
Weight
< (SP2 – SP6)
Upper Limit NG: ON when (SP2 + SP5) <
Weight
Near Zero : F1-23 set value ≥
Weight
≥ 0 range output
4. Status lamp in the front panel is operated in the same manner as the RELAY output