3. Installation & Connection
3-1. Loadcell Connection
Connect the load cell connector to the load cell port which is in the backside of the
indicator.
* Connection method
Note 1. In case of 4 wires load cell, connect EX+ with SEN+, and connect EX- with SEN-.
Note 2. Wire color can be different depending on the load cell’s manufacturer or it’s model.
* Relationship between the load cell output and input sensitivity.
The input sensitivity of this product is maximum 0.2 uV/digit or more.
The following equation should be satisfied upon the system design.
Pin
No
Pin
Function
Color
1
EXC+
RED
2
EXC-
WHITE
3
SIG+
GREEN
4
SIG-
BLUE
5
SHIELD
6
SEN+
BROWN
7
SEN-
BLACK