CI-100A Indicator Operation & Service Manual
────────────────────────────────────────────────────────
CAS-USA
Corporation
www.cas-usa.com
29
the weight that is over
USER-OTHER-NLD.RNG (
zero 'dead' band ) can be held. To exit
HOLD
mode,
press
HOLD
key again. If the time of scale being unstable is more than
USER-HOLD-STB.TIM
,
STB.ER
will be shown, press
TARE
key to start averaging again, or press
HOLD
key to exit.
9.3.4
Average HOLD
:
When
USER-HOLD-HLD.MOD
is set to
AVERAG
, the hold mode is average hold mode. After entering
this working mode, scale will freeze and display number if scale is stable. If scale is not stable, but the
variation is less than
USER-HOLD-HLD.RNG,
scale will average data in
USER-HOLD-AVG.TIM
, then
freeze and display the number. Only the weight that is over
USER-OTHER-NLD.RNG
can be frozen.
Scale will exit HOLD mode according to the setting of
USER-HOLD-HLD.TIM
. If the time of scale
variation being over
USER-OTHER-NLD.RNG
is more than
USER-HOLD-STB.TIM
,
STB.ER
will be
shown, press
TARE
to start averaging again, or press
HOLD
key to exit.
9.3.5
Auto HOLD:
When
USER-HOLD-HLD.MOD
is set to
AUTO
, the hold mode is auto hold mode--- different subjects
can be weighed one after another without pressing any buttons. After entering this working mode, scale
will freeze and display number if scale is stable. If scale is not stable, but the variation is less than
USER-HOLD-HLD.RNG,
scale will average data in
USER-HOLD-AVG.TIM
, then freeze and display the
number. Only the weight that is over
USER-OTHER-NLD.RNG
can be frozen. If the held weight is
moved away, and a new load put on the scale, scale will automatically hold new number of load. Scale
will exit HOLD mode according to the setting of
USER-HOLD-HLD.TIM
. If the time of scale variation
being over
USER-OTHER-NLD.RNG
is more than
USER-HOLD-STB.TIM
,
STB.ER
will be shown,
press
TARE
to start averaging again, or press
HOLD
key to exit.
9.4 In Positive or Negative Peak HOLD mode, the red HOLD (for LED version) or PEAK and HOLD (for LCD
version) announciator will be lighted, in other HOLD mode, green HOLD (for LED version) or HOLD (for
LCD version) announciator will be lighted. When HOLD announciator flash, the displayed number is live,
When HOLD announciator become steady, the displayed number is frozen.
10. Data Compare Function
10.1 Data compare function can be used in normal weighing mode, counting mode and percent weighing
mode, and call it as Check Weight, Check Counts and Check Percentage. When this function is enabled,
you can set a higher and a lower limitation of weight, counts or percentage independently, and these
limitation can be saved permanently. Then, the current data of weight , counts or percentage will be
compared with the setting limitation, and corresponding annunciator will be lighted.
10.2 To make data compare function be available,
CONFIG-FUNC-COMPAR
menu item should set to
YES
,
and high and low limitation should be set correctly according to following steps:
10.3 In normal weighing mode, counting mode or percent weighing mode, Press down
DATA
key more than
3s to enter input compare data of high and low mode.
10.4 After
HIGH
being shown, last setting data of high will be displayed, use
,
ACC
,
UNIT
key to input
new number of high and press
TARE
key to confirm. Annunciator of
Hi
will be shown in this step. Press
ZERO
key to exit and back to original working mode.