CI-100A Indicator Operation & Service Manual
────────────────────────────────────────────────────────
CAS-USA
Corporation
www.cas-usa.com
48
CACU.ER
Internal calculation overflow
Adjust the value of the PWT or PCT
STB.ER
USER-HOLD-STB.TIM is too short,
USER-HOLD-HLD.RNG is too small,
other failure
Set USER-HOLD-STB.TIM longer,
or set USER-HOLD-HLD.RNG
bigger. Service required
Not turn on.
Power cord not plugged in or properly
connected. Power outlet not supplying
electricity. Battery discharged. Other
failure.
Check power cord connections.
Make sure power cord is plugged
into the power outlet. Check power
source. Replace batteries. Service
required.
Cannot zero the
display or will not zero
when turned on.
Load on scale exceeds allowable limits.
Load on scale is not stable.
Load cell damage.
Remove load on scale.
Wait for load to become stable.
Service required.
Cannot display weight
in desired weighing
unit.
Unit not set to enable, or d
≥
5oz,when
unit is lb:oz
Enable unit in CONFIG-UNITS
Battery symbol is
empty or Lo.bAt is
shown
Batteries are discharged.
Charge batteries