CI-100A Indicator Operation & Service Manual
────────────────────────────────────────────────────────
CAS-USA
Corporation
www.cas-usa.com
32
weight for third point. Put corresponding weight (more than 10%FS weight, and larger than the weight
used on
CAL.P2
) onto scale. Next operation is same as what in step12.5.2.
12.5.7 When the stable and reasonable data corresponding to the standard weight has been gotten. The
indicator will automatically go to Step12.8. Otherwise, it will display “
CAL.Er
” and return back.
12.6 When
GEO
is selected, press
TARE
key to confirm to enter Geographical Adjustment
12.6.1 When “
CODE
” is shown, use
or
ACC
key to select geographical position code (refer step12.6.2)
or input user local gravity value directly (refer step12.6.3).
12.6.2 When
CODE
is selected, select the position code of scale being used (00-70) according to the
elevation and latitude from Table12-1 by using
PRINT, ACC, UNIT
keys. Press
TARE
key to confirm.
12.6.3 When
GRAVT
is selected, Use
PRINT, ACC, UNIT
keys to input the gravity value of the position that
scale is used (9.76183-9.99999). Press
TARE
key to confirm.
NOTE: Only an authorized manufacturer’s representative or certified verification personnel
may make these changes. Changing the geographical setting alters the calibration
values !!!
12.7 When
INPUT
is selected, press
TARE
key to confirm to enter Input calibration parameters value that
were got before, or view current calibration parameters value.
12.7.1 All parameters about calibration are divided to 18 pages to be displayed on LCD by “nn:xxxx” format
(“nn” is a decimal number of page, “xxxx” is an hexadecimal value of parameter, e.g. 02:85E2).
01-02 pages: zero code;
03-04 pages: standard weight of CAL.P1;
05-06 pages: codes of CAL.P1;
07-08 pages: standard weight of CAL.P2;
09-10 pages: codes of CAL.P2;
11-12 pages: full capacity net code;
13-14 pages: the coefficient of weight fine-tune;
15-16 pages: gravity value of calibration location;
17-18 pages: gravity value of the location that the scale is used at.
12.7.2 When no digit blink on LCD, that means calibration parameters value are being viewed, and use
UNIT
key to view next page, use
ZERO
key to exit.
12.7.3 When parameters value are being viewed, use
DATA
key to prepare to modify, When first digit is
blinked, that means the value is being modified, and you can use
UNIT
key to make next digit flash (if
current flashing position is the last one, next page value will be shown), use
PRINT, ACC, UNIT
key to
input number, use
TARE
key to confirm.
12.7.4 In this mode, press down
key more than three seconds, these parameters will be sent out from
COM1, the print out format is <LF>nn:xxxx<CR>; there’re totally eighteen lines.
12.8 After the indicator gets all need data, it will calculate and store all calibration parameters into EEPROM,
or after finishing calibration works, it will display
CAL.End
. At last, it will re-start and go back to original
mode.
Содержание CI-100A
Страница 1: ......
Страница 2: ...CI 100A Indicator Operation Service Manual CAS USA Corporation www cas usa com ...