CI-100A Indicator Operation & Service Manual
────────────────────────────────────────────────────────
CAS-USA
Corporation
www.cas-usa.com
27
percentage of samples and Press
TARE
key to confirm. If the calculated unit-percentage-weight is
less than 0.5d, the indicator will display
Pct.Er
and go back to percent weighing mode, otherwise,
after the reasonable unit-percentage-weight being got, the scale will go back to percent weighing
mode. The got unit-percentage-weight can be saved after the power off and can be used next time.
7.6
Check Percent (percentage compare)
in Percent weighing mode:
7.6.1
To make percentage compare function available,
CONFIG-FUNC-COMPAR
menu item should set to
YES
, and high and low limitation of percentage should be set correctly according to following steps:
7.6.2
In percent weighing mode, Press
DATA
key more than 3s to input compare data of high and low.
7.6.3 After
HIGH
being shown, 000000 will be displayed, use
,
ACC
,
UNIT
key to input high
percentage number and press
TARE
key to confirm. Annunciator of
Hi
will be shown in this step.
Press
ZERO
key to exit and back to percent weighing mode.
7.6.4 After
Low
being shown, 000000 will be displayed, use
,
ACC
,
UNIT
key to input low
percentage number and press
TARE
key to confirm. Annunciator of
Lo
will be shown in this step.
Press
ZERO
key to exit and back to counting mode.
NOTE:
If High number is 0 or is equal or less than low number, the comparison will be disabled.
7.6.5
After a reasonable limitation is set and compare is be active, one of the annunciator,
HI, OK, LO
will
be lighted, and the beeper will sound according to its setting in
USER-BEEP
.
8. BMI Working Mode
8.1
To make BMI working Mode be available,
CONFIG-FUNC-BMI
menu item should be set to
YES
and
factory setting should be also enable this function
8.2
To enter BMI Working mode:
8.2.1 When
CONFIG-FUNC-ACCUMU= Yes:
If In normal weighing mode, percent weighing mode, or
counting mode, long press
FUNC
key, one of
WEIGH/COUNT/PERCEN
will be shown, use
ACC
or
key to select
BMI
, then press
TARE
to confirm go to BMI mode.
8.2.2 When
CONFIG-FUNC-ACCUMU= NO:
press
BMI
(
ACC
) key, go to BMI mode.
8.3
After scale go to this mode, “CM.xxx” (means: last input height is xxx cm) or “IN.xx.x” (means: last
input height is xx.x inch) will be displayed, and to wait for input height: (1)to change height unit to cm
or inch by pressing
DATA
key; (2) to change height number by using
PRINT, ACC(BMI)
keys
(3)Press and hold down
PRINT or ACC(BMI)
key will increase or decrease number fast, (4) Press
TARE
key to confirm the input. Press
ON/OFF
key to exit input data mode and back to BMI working
mode. The range of height is 50-250cm(19.7-98.4inch) and default is 170cm(66.9inch)
8.4
In this mode, when BMI number is shown (BMI annunciator is on also), or weight number is shown
(BMI and kg or lb annunciator are on), Press
ACC
key to select weight or BMI number to be
displayed, when weight is displayed, the weight unit can be selected by pressing
UNIT
key, and BMI
and weight unit will be displayed at same time.
8.5
In this mode, when current net weight is less than
NLD.RNG
, the indicator will go to display weight
number
if
CONFIG-FUNC-ACCUMU
= No; or the indicator will back to original working mode if
CONFIG-FUNC-ACCUMU
= Yes.