23.6.1.5 Command:
U<CR>
(55h 0dh)
Response: Changes units of measure (simulate
UNIT
key) and return scale status with
new
units,
The new measure unit should be allowed to use
<LF> U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>
23.6.1.6 Command:
L<CR>
(4ch 0dh)
Response: If Hold function can be activated, it will enable/disable hold function (simulate
HOLD
key),
and returns scale status.
<LF> H
1
H
2
H
3
H
4
<CR><ETX>
23.6.1.7 Command:
X<CR>
(58h 0dh)
Response: power off the scale, just like press down the
ON/OFF
key to turn off the scale.
23.6.1.8 Command:
all others
Response: Unrecognized command
<LF>? <CR><ETX>
23.6.2 Summary of Command and Response:
Command
Response
ASCII
HEX
W<CR>
57 0d
Read scale weight:
<LF>^^^^^^^^U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---over capacity
<LF>________U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---under capacity
<LF>-------- U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---zero-point error
<LF><p>W
1
W
2
W
3
W
4
W
5
<dp>W
6
U
1
U
2
U
3
U
4
U
5
<CR><LF>H
1
H
2
H
3
H
4
<CR><ETX>
---normal data
S<CR>
53 0d
<LF> H
1
H
2
H
3
H
4
<CR><ETX>; read scale status
Z<CR>
5a 0d
<LF> H
1
H
2
H
3
H
4
<CR><ETX> ; simulate ZERO key
T<CR>
54 0d
<LF> H
1
H
2
H
3
H
4
<CR><ETX> ; simulate TARE key
U<CR>
55 0d
<LF> U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>; simulate UNIT key
L<CR>
4c 0d
<LF> H
1
H
2
H
3
H
4
<CR><ETX>; simulate HOLD key
X<CR>
58 0d
power off the scale, simulate OFF key
others
<LF>? <CR><ETX>