CI-100A Indicator Operation & Service Manual
────────────────────────────────────────────────────────
CAS-USA
Corporation
www.cas-usa.com
18
4.4 CAL Submenu:
CAL
SUBMENU1 SUBMENU2
OPTION
REMARK
CAL.ON
CAL.OFF
seal switch is on or off
ZERO
only do zero point calibration, then go to CAL.END to
end
LINE CAL.P0
Linear calibration point0: do zero point
calibration, this point can't be omitted.
CAL.P1
Linear calibration point1: do first weight point
calibration, this point can't be omitted and standard
weight must be over 10%FS.
END.Y
YES
End calibration? YES=go to CAL.END to end; NO=go
to do next point calibration
NO
CAL.P2
Linear calibration point2: do second weight point
calibration, standard weight must be over 10%FS
and be larger than it in CAL.P1, this point can be
omitted.
END.Y
YES
End calibration? YES=go to CAL.END to end; NO=go
to do next point calibration
NO
CAL.P3
Linear calibration point3: do third weight point
calibration, standard weight must be over 10%FS
and be larger than it in CAL.P2, this point can be
omitted.
GEO
CODE
00-70
Selection of Geographical Position Code 00-70,refer
TABLE12-1
GRAVT
9.76183
-9.99999
Input Gravity of User Location by keyboard
INPUT
Input or view calibration parameters value
CAL.END
calibration end and restart
NOTE:
The details can be referred in section
“12.CALIBRATION”