Section 7: Technical Data
Survivalink AED Models 9200/9210 Waveform
AAMI DF-2 (1996), Section 4.3.4.3, other waveforms
Low Current Survivalink AED Models 9200/9210 Waveform
(all values are typical)
High Current Survivalink AED Models 9200/9210 Waveform
(all values are typical
)
Phase 1
Phase 2
Patient’s Impedance
(Ohms)
Voltage
(Volts)
Duration
(ms)
Voltage
(Volts)
Duration
(ms)
Energy
(Joules)
25
1570
3.3
825
3.2
180-250
50
1600
4.5
1031
3.2
170-220
75
1620
5.8
1111
3.2
150-210
100
1630
7.0
1158
3.2
140-190
125
1650
8.3
1193
3.2
140-190
Phase 1
Phase 2
Patient’s Impedance
(Ohms)
Voltage
(Volts)
Duration
(ms)
Voltage
(Volts)
Duration
(ms)
Energy
(Joules)
25
1890
3.3
993
3.2
270-360
50
1920
4.5
1238
3.2
240-320
-1500
-1000
-500
0
500
1000
1500
2000
2500
0
5
10
15
20
Time (ms)
Output
Voltage
(V)