300282-002 Rev. E0
©2002 Cardiac Science, Inc.
Page 41
Section 4: Rescue
Audible alerts
“Two-Tone Beep” occurs after insert-
ing an optional Rescue data card or a
MDLink options card into the Surviv-
alink AED’s card slot, with the lid
open. Also occurs in 15-second inter-
vals during CPR when enabled by the
MDLink software program.
“Warble Beep” occurs when the Sur-
vivalink AED requires maintenance.
Press flashing but-
ton to continue
rescue
PRESS BUTTON
TO CONTINUE
RESCUE
When you open the Survivalink AED
lid and the electrodes are attached to
the patient and connected to the Surviv-
alink AED. The YELLOW “Resume”
indicator flashes and the phrase will
repeat until you press the “Rescue/
Resume” button.
Continue CPR
CPR X:XX
Phrase repeats in 15-second intervals
during CPR mode. You can enable the
“
Continue CPR
”
prompt using the
MDLink software program.
Asystole
ASYSTOLE
Occurs when the Survivalink AED
detects “Asystole” during ECG analy-
sis. You can enable the “Asystole”
prompt using the MDLink software
program.
Service required
SERVICE REQUIRED
Occurs when the self-tests determine
that the Survivalink AED is not func-
tioning properly.
The RED “Service” indicator will illu-
minate and
“
Service required
”
will
repeat until you close the lid. After
closing the lid, a “warble beep” will be
heard until the battery is removed or
becomes completely depleted.
Voice Prompt
Optional Text Display
9200/9210 Models only
When You Will Hear the Prompt